Freedom KD2-00-KNEE2-KT Instructions For Use Manual Download Page 16

b. With equal weight on each limb, adjust the height of the prosthesis as necessary.
c.

Ensure the trochanter-knee-ankle (TKA) alignment follows the recommendations of inherent stability.

d. Ensure the transverse rotation of the knee and foot is appropriate.
e. Instruct the user  to sit  in  a  chair. Adjust the height of the knee joint axis  to match  the contralateral  limb  as 

closely as possible.

!

To prevent risk of injury to the user, perform the static alignment, dynamic alignment, and set-up between 

!

parallel bars.

Dynamic Alignment

Prior to Set-Up, allow the user to adjust to the prosthesis:
a. Instruct  the  user  to  take a  lunge step  with  the  prosthesis.  The lunge motion  will  allow  the user  to  feel  the 

stance flexion resistance

 and develop confidence.  Repeat as necessary.

b. Instruct  the  user  to  carefully  ambulate. Teach  the user  to flex  the ipsilateral  hip  extensor  muscles at  initial 

contact to stabilize the knee joint. 

c.

Train the user to load the prosthetic toe to initiate the swing phase transition.

d. Train the user to take steps of equal length.
e. Adjust the alignment in the transverse, coronal, and sagittal planes as necessary. 
f.

Train the user to sit in a chair using the 

stance flexion resistance

.

Set-Up

While  the Plié®  2.0  MPC Knee may  function  adequately at the default settings, both  the computer and manual 
hydraulic  settings should  be optimized  for  each  user.  Follow  the  five steps  below  to  set-up  the Plié®  2.0  MPC 
Knee.

1.  Establish Wireless Connection

To  optimize the  microprocessor,  communication  must  first  be established between  the  programming  computer 
and the user

ʼ

s Plié® 2.0 MPC Knee:

a. Open the Plié 2.0 Control software (Start>All Programs>Plié>Plié Control). 
b. Insert the provided USB wireless adapter in any available USB port.
c.

Remove and  re-insert the battery  in  the  knee.  Battery  insertion  activates  the knee  wireless  module for  five 
minutes.

d. Select the 

Search

 icon.  

e. After  the knee is  found,  select  the 

Connect

 icon  to establish  a  wireless  connection.    If  the 

Connect

 icon 

changes to a 

Disconnect

 icon, the wireless connection is verified.

!

If  the  software  and  knee  do  not  appear  to  be  communicating,  select  the 

Wireless  Sync

 

icon  to  re-

!

establish the wireless connection.  

!

The green LED inside the knee frame only remains illuminated when the a wireless connection has been 

!

established with a computer.

2.  Plié 2.0 Setup Wizard

Open the Plié 2.0 Setup Wizard (Plié Control>Setting Wizard). The 

Welcome

 window will appear.  

a. If you are programming a knee for the first time, select 

Next

.

b. In the 

Setup and Calibration

 window, select 

Setup Mode

.  

The  Setup  mode  should  only  be  enabled  when  setting  up  a  knee.  The  advanced  Stumble  Recovery 
Parameters (Levels I & II) are not active and the stumble recovery function is limited in the Setup mode.

c.

As the user unweights the prosthesis, select the 

Calibrate

 icon  to calibrate the load sensor. Instruct the user 

to ambulate  or  weight  the prosthesis  in  a  fully  extended  position  to  calibrate the angle sensor.  Select 

OK

 

when finished.

Tip

Tip

Plié® 2.0 MPC Knee Instructions for Use

 

(R-720-109 Rev. E), English

!

!

!

!

!

        16

Summary of Contents for KD2-00-KNEE2-KT

Page 1: ...Plié 2 0 MPC Knee Instructions for Use R 720 109 Rev E English 1 INSTRUCTIONS FOR USE ...

Page 2: ...0 Plié Gait Lab 13 Alignment 14 Bench Alignment 15 Static Alignment 15 Dynamic Alignment 16 Set Up 16 1 Establish Wireless Connection 16 2 Plié 2 0 Setup Wizard 16 3 Adjust Manual Hydraulic Settings 17 4 Optimize Microprocessor Settings 17 5 Document 18 Technical Support 18 Warranty 18 Service and Repair 18 Regulatory Compliance Disclosures and Classification 19 Wireless Operation 19 IEC EN 60601 ...

Page 3: ...Plié 2 0 MPC Knee Instructions for Use R 720 109 Rev E English 3 H D C A B E F G Plié 2 0 MPC Knee images not to scale ...

Page 4: ...Plié 2 0 MPC Knee Instructions for Use R 720 109 Rev E English 4 N I J K L M O Recommended Foot Components P Q R ...

Page 5: ...e KS2 00 UMAN1 00 1 Plié Lower Shank Cosmetic Foam KS2 00 COVR 00 Highlander foot component FS3 Runway foot component RS2 or RS2 K1 Renegade foot component RS1 or RS6 Appropriate Use The Plié 2 0 MPC Knee should only be used by individuals with lower limb loss including knee disarticulation hip disarticulation transfemoral amputees and or individuals with congenital anomalies The Plié 2 0 MPC Knee...

Page 6: ...unction Use only the batteries and battery charger intended for use with the Plié 2 0 MPC Knee Extreme temperatures will affect the performance of the Plié 2 0 MPC Knee Allow sufficient time for the knee to return to temperatures within the specified operating temperature range before use High elevations and cold temperatures will affect the knee internal air pressure In these conditions the user ...

Page 7: ...ricate the shaft with a few drops of 3 in1 oil All purchased serviced and repaired Plié 2 0 MPC Knees are sent to customers with the following default hydraulic cylinder manual settings swing flexion 30 psi swing extension 0 stance flexion 5 Additionally the computer settings are sent to customers with default settings The prosthetist should adjust the manual and computer settings to meet the need...

Page 8: ...Hydraulic cylinder default manual settings swing flexion 30 psi swing extension 0 stance flexion 5 Patents USPO 5 190 126 6 978 872 and 7 655 050 Other patents pending Battery and Battery Charger Specifications Battery type lithium ion Battery voltage 3 7 VDC Battery capacity 710 mAh Battery life 24 hours depending on use Charger input voltage 12 VDC AC power adapter input voltage 100 240 VAC 50 6...

Page 9: ...ed to 90 capacity the Charge indicator LED will flash red and green When the battery is fully charged the Charge indicator LED will illuminate green Charging of a fully discharged battery takes approximately two hours Use only approved batteries and battery charger battery part AG BATT XX and battery charger part AG CHRG XX where XX specifies current revision number Use of unapproved batteries and...

Page 10: ...isconnect Select the Connect Disconnect to establish or discontinue a wireless connection 4 Toe Threshold T1 The Toe Threshold determines how much force the user must load the toe to initiate swing phase during ambulation A higher value requires more toe loading by the user to initiate swing phase The Toe Threshold may be adjusted by moving the slider or typing a numerical value in the text box an...

Page 11: ...ssion of the Plié 2 0 Setup Wizard see section titled Set Up 10 More Params Select More Params to open the auxiliary Additional Parameters window for the advanced Stumble Recovery parameters Figure D 11 Data Monitor Select Data Monitor to open an auxiliary window for diagnostic use only 12 Live Data Select Live Data to view the auxiliary Plié Real Time Data Display window for viewing data as the u...

Page 12: ...ime Data Display is illustrated in Figure C The Plié 2 0 Real Time Data Display is useful for verifying calibration of the sensors If the user unweights the knee the Moment trace should have zero value If the knee is fully extended the Angle trace should have zero value Additional Parameters The Additional Parameters window provides access to advanced stumble recovery parameters via two tabs Level...

Page 13: ...lect the optimal Toe Threshold Swing Transition and advanced Stumble Recovery Parameters Levels I II Plié Gait Lab The Plié Gait Lab software allows the prosthetist access to the userʼs gait data as viewed by the microprocessor The Plié Gait Lab may be used as a tool for comparing the performance of various prosthetic feet identifying gait anomalies simulating the microprocessor algorithm ensuring...

Page 14: ...he prosthesis the prosthetist should measure the userʼs hip joint range of motion ROM on the prosthetic side to determine if the user has a hip flexion contracture Figure E If present the userʼs hip flexion contracture should be accommodated by attaching the socket to the prosthesis with an appropriate amount of flexion Figure F Failure to sufficiently accommodate a hip flexion contracture may com...

Page 15: ...guidelines An inherently unstable trochanter knee ankle TKA alignment may cause a rapid extension moment at the knee joint after midstance The non adjustable stance extension resistance may be insufficient to dampen the extension moment if the user does not exert sufficient voluntary control and or if the alignment recommendations are not implemented The recommended inherently stable alignment of ...

Page 16: ...e 1 Establish Wireless Connection To optimize the microprocessor communication must first be established between the programming computer and the userʼs Plié 2 0 MPC Knee a Open the Plié 2 0 Control software Start All Programs Plié Plié Control b Insert the provided USB wireless adapter in any available USB port c Remove and re insert the battery in the knee Battery insertion activates the knee wi...

Page 17: ...or the remainder of the knee set up proceed with the Plié 2 0 Setup Wizard Plié Control Setting Wizard a Instruct the user to ambulate with variable cadences including fastest and slowest walking speeds until 20 steps are shown in the Steps box The Plié 2 0 Setup Wizard will select optimal T1 T2 values and transfer them to the knee The suggested values in the Set T1 and T2 boxes should have minima...

Page 18: ...n a secure computer To save select Save Settings for the File menu Plié Control File Save Settings The userʼs saved settings can be transferred to a loaner knee However due to continuous design improvements the settings may not be optimal If problems are experienced with a loaner knee simply follow the steps provided in this Set Up section Technical Support For Technical Support please contact Fre...

Page 19: ...mable anesthetic mixture with air or with oxygen or nitrous oxide Directive 2002 96 EC on Waste Electronic Electrical Equipment WEEE We have marked the Plié 2 0 MPC Knee with the crossed out wheeled bin as instructed by Article 10 3 of WEEE per Annex IV We include the following instruction required by Article 10 of WEEE for the Prosthetist to convey to the End User We instruct Users of the Plié 2 ...

Reviews: