Freedom Innovations Plie3 MPC Knee Instructions For Use Manual Download Page 7

R-720-190 Rev. 

E

 English

4.1   Alignment 

Careful attention to the alignment of the socket in relation to the Plié® 3 MPC Knee and the 

prosthetic foot is essential for a successful user outcome. The prosthesis alignment should 

account for the range of motion (ROM), voluntary control, and balance of individual users. 

Proper alignment and user voluntary control are essential to the optimal function of the 

prosthesis. 
Prior to assembly of the prosthesis, the prosthetist should measure the patient

’s hip joint range 

of motion (ROM) on the prosthetic side to determine if the user has a hip flexion contracture 

(Figure 1

). If present, the user’s hip flexion contracture should be

 accommodated by attaching 

the socket to the prosthesis with an appropriate amount of flexion (Figure 2). Failure to 

sufficiently accommodate a hip flexion contracture can compromise the patient

’s function 

during standing and ambulation (Figures 3 and 4).   

During standing, an unaccommodated hip flexion contracture may prevent the patient 

from standing straight, compromising the patient

’s balance (Figure 

5). Additionally, an 

unaccommodated hip flexion contracture can cause excessive lumbar lordosis, 

compromising the structural integrity of the patient

’s spinal column (Figure 

4). 

During ambulation, an unaccommodated hip flexion contracture can cause an 

excessively asymmetrical gait pattern. 

4.2

 

 Bench Alignment 

The Plié® 3 MPC Knee should be in a fully extended position during stance phase for level 
ground ambulation.  An inherently stable trochanter-knee-ankle (TKA) alignment is essential to 
a successful user outcome with the Plié MPC Knee.  The referenced trochanter on the lateral 
socket should be placed on or up to 5mm anterior to the knee joint axis (Figure 5).   

An inherently unstable trochanter-knee-ankle (TKA) alignment can cause a rapid 

extension moment at the knee joint after mid stance. The non-adjustable stance 

extension resistance could be insufficient to dampen the extension moment if the 

user does not exert sufficient voluntary control and/or if the alignment 

recommendations are not implemented. 

The recommended inherently stable alignment of the Plié 3 MPC Knee may differ 

from other knee joints. Consequently, if a new socket is not fabricated for use with a 

Plié 3 MPC Knee, the socket attachment component may require re-lamination to the 

socket to achieve the recommended alignment 

4.3

 

Static Alignment 

4.3.1

 

Instruct the user to stand between parallel bars.

 

4.3.2

 

With equal weight on each limb, adjust the height of the prosthesis as necessary.

 

4.3.3

 

Ensure

 

the trochanter-knee-ankle (TKA) alignment follows the recommendations of 

inherent stability.

 

Summary of Contents for Plie3 MPC Knee

Page 1: ...R 720 190 Rev D English 1 INSTRUCTIONS FOR USE MICROPROCESSOR KNEE...

Page 2: ...tions 5 Battery and Battery Cap 4 Alignment and Setup 6 4 1 Alignment 4 2 Bench Alignment 4 3 Static Alignment 4 4 Dynamic Alignment 4 5 Pli Control 6 4 6 Pli 3 Setup 4 7 Finishing Options 4 8 Technic...

Page 3: ...Cover 2 Pli 3 MPC Knee Wireless USB Adapter 3 Pli 3 MPC Knee Lithium Ion Battery Case 4 Pli 3 MPC Knee Software CD V6 5 Pli 3 MPC Knee Litium Ion Battery 2 pc 6 Pli 3 MPC Knee Battery Charger with Ca...

Page 4: ...r individuals with lower limb loss including transfemoral amputees knee disarticulation amputees hip disarticulation amputees individuals with congenital lower limb abnormalities The Pli 3 MPC Knee is...

Page 5: ...1 3 Insert a charged battery when the Pli 3 MPC Knee low battery indicator is visible symbolized by red blinking light visible on the left side of the knee viewed from posterior prior to complete powe...

Page 6: ...nment and Setup Follow the Prosthe t Setup Guide to properly align and adjust the device Two installa on programs Pli Control and Pli Gait Lab are provided to op ize the computer se ngs for each pa en...

Page 7: ...t s spinal column Figure 4 During ambulation an unaccommodated hip flexion contracture can cause an excessively asymmetrical gait pattern 4 2 Bench Alignment The Pli 3 MPC Knee should be in a fully ex...

Page 8: ...Figure 5 4 4 Dynamic Alignment 4 4 1 Instruct the user to take a lunge step with the prosthesis The lunge motion will allow the user to feel the Stance Flexion resistance and develop confidence Repeat...

Page 9: ...5 Warranty and Service The Pli 3 MPC Knee purchase includes a 36 month warranty covering all manufacturer defects effective only if the product is used according to manufacturer recommendations The ba...

Page 10: ...3 psi to 15 4 psi Water and Dust Resistance IP67 Totally protected against dust 6 and Protected against the effect of immersion between 15cm and 1m 7 6 4 CE Conformity The electronic of Pli 3 MPC Kne...

Reviews: