Safety Instructions

                     

For your protection

• 

Read the operating instructions in full before starting up

   and follow the safety instructions when in use.

• Ensure that only trained staff work with the device.

• Socket must be earthed (protective ground contact)

• 

Caution-Magnetism!

 Effects of the magnetic field have to be

   taken in to account (e.g. data storage media, cardiac

   pacemakers…).

• Wear your personal protective equipment in accordance with

   the hazard category of the media to be processed. Otherwise

   there is a risk from:

   - Splashing and evaporation of liquids

   - Ejection of parts

   - Release of toxic or combustible gases

• Set up the device in a spacious area on an even, stable, clean,

   nonslip, dry and fireproof surface.

• Don’t use damaged components.

• Gradually increase the speed and reduce the speed if

   - the medium splashes out of the container because the

     speed is too high

   - the machine is not running smoothly

   - the container moves on the heating plate

• Beware of hazards due to:

   - glass breakage

   - incorrect container use

   - overfilling of media

   - unsafe condition of container

• Don’t use the machine in explosive atmospheres with

   hazardous substances.

• Process pathogenic materials only in closed containers under

  a suitable extractor hood.

2

Summary of Contents for MI0102008

Page 1: ...Operating Instructions 5 Inch LED Digital Magnetic Stirrer with Timer MI0102008...

Page 2: ...Contents Safety Instructions Inspection Control Trial run Timing Stirring Faults Error codes Maintenance Standards and regulations Specifications Warranty 2 3 4 6 6 7 7 7 7 8 9 10 1...

Page 3: ...from Splashing and evaporation of liquids Ejection of parts Release of toxic or combustible gases Set up the device in a spacious area on an even stable clean nonslip dry and fireproof surface Don t...

Page 4: ...odic system at temperatures above 300 C 400 C Only elementary fluorine chlorotrifluoride and alkali metals attack it halogenated hydrocarbons have a reversible swelling effect The voltage stated on th...

Page 5: ...Contents of package 4 Qty 1 1 1 1 1 Items Main unit Power cable Screwdriver Magnetic stirring bar User Manual Control...

Page 6: ...e speed by rotating the knob Start stop stirring by pressing the knob When timer function starts it displays the remaining stirring time It displays OFF when time is off When stirring it displays the...

Page 7: ...the remaining stirring time Timer function is switched on off by pressing the timer setting If the above operations are normal the device is good to run If these operations are abnormal the device may...

Page 8: ...en restart Actual speed can t reach the set value Media in high viscosity may probably cause abnormal speed reduction of the motor Error codes 7 If the instrument still can t work properly please cont...

Page 9: ...cohol water containing tenside water containing tenside Standards and regulations Construction in accordance with the following safety standards EN 61010 1 UL 3101 1 CAN CSA C22 2 1010 1 EN 61010 2 10...

Page 10: ...range rpm Speed display Working plate material Dimension of working plate mm Time setting range Timing display Dimensions W x D x H mm Weight kg Permissible ambient temperature C Permissible relative...

Page 11: ...nly to the original purchaser It doesn t cover any worn out parts nor apply to any damage by improper use insufficient care or maintenance not carried out in accordance with the instructions in this o...