2. Carefully loosen and pull out the old filter seal. Press gently
down the new filter seal.
Note: Check carefully the correct positioning of the new
filter seals.
5.4
Change of the lower thimble
holder
1. Loosen the old thimble holder by gently easing and pulling it
upwards.
2. Press down the new thimble holder and check that it fits
correctly.
5.5
Change of the vacuum tube
1. Remove the plug from the wall socket.
2. Unscrew the instrument panel with the heating control knob.
3. Remove the white plate on the back of the hydrolyzing unit
according to paragraph 5.1 Pos. 1- 4.
4. Remove the plastic strap on the neoprene tube that is situated
to the far right on the connection tube.
5. Unscrew the nipple on the back of the hydrolyzing unit that is
marked "vacuum".
6. Pull out the tube with the nipple through the instrument hole.
7. Cut the new vacuum tube to 450 mm and assemble the nipple
from the old tube.
8. Assemble the rest in reverse order.
5.6
Change of fuses
1. The 2 pole fuse holder is located under the mains cable. Push
the middle button and change the fuses.
1047
Soxtec
System
Hydrolyzing
Unit,
User
M
anual,
R
ev.
2.2,
Part
no
1000
3315,
Foss
Tecator
AB,
Sweden
12
1047012A
1
2
3
4
5
VACCUM
WATER
DRAIN
!
Fig. 11
Pos 2: 2 pole fuse holder and mains