c.
How to deal with the problem of LED light above screen turns red?
@ This indicates the electric capacity of main battery is in a weak position. Please
charge or replace the battery as soon as possible to keep the work normally
executed.
d.
How to solve the problem that a blue page shows up while the testing page
suddenly disappears as using?
@ This is caused by wrong execution of programs. Please push F1+F4 button to
restart the Tester, or execute the step of power off / power on.
e.
No messages can be found when scanning the base stations :
@ This often occurs due to a repeated scanning within a short period. It takes 10
seconds for the Tester to make a complete scan, but a hurried scan could
probably enforce the Tester to continue the next scan without ending
previous one, then a programming error might happen. Please restart the
Tester by the step of power off / power on.
15.
Trouble
Shooting