Position light for oral cavity illumination using handles for focal length and direction.
•
Turning On. Toggle switch (Figure
A
) left.
= On = Off
•
Automatic On/Off. Lift light flex arm above and below 0˚.
(To deactivate contact your service technician.)
•
Adjusting light intensity. Use knob (Figure
B
) to adjust intensity.
•
Adjusting color temperature. Use knob (Figure
C
)
to adjust color temperature.
Removing & Replacing Front Shield
CAUTION: Never touch the LED reflectors or lamps. If necessary, clean only with low pressure air.
MISE EN GARDE: Ne touchez jamais les réflecteurs ou les lumières LED. Si nécessaire, nettoyer uniquement avec de l’air à basse pression.
1. Turn the 1/4 turn fasteners on front of shield clockwise. (Figure
D
)
2. Pull shield out with fasteners. See (Figure
E
)
3. Reverse to reinstall. See (Figure
F
)
Removing & Sterilizing Handles
1. Press in the locking tab. (Figure
G
)
2. To completely unhook, wiggle and pull handle outward while
pressing in locking tab.
3. Autoclave @ 250° F (121° C) for 10 minutes or 273° F (134° C)
for 4 minutes.
4. Reverse steps 1 and 2 to reinstall.
Cleaning Front Shield & Reflector
17UJ
A
B
C
D
E
F
G
H
I
CAUTION: Abrasive or detergent materials such as alcohol, trichloroethylene, gasoline (petrol), turpentine or the like should not be used on any part of
the light.
MISE EN GARDE: Les matériaux ou détergents abrasifs comme l’alcool, le trichloréthylène, l’essence, la térébenthine et toute autre substance semblable ne
devraient pas être utilisés sur aucune partie de la lampe.
1. Remove front shield. (Figures
D & E
)
2. Clean the front shield with a soft cloth moistened with mild soapy water. Using chemical disinfectants can damage components. The front
shield is not autoclaveable.
3. If needed, clean reflector with low pressure air only.
LED Light
Operating Light
Light Friction Adjustment Screw
1. Looking at the front of the light, remove the right side decorative trim from the back of the light by pulling it up with your fingers. This
will reveal the friction adjustment screw. (Figure
H
)
2. Using a 1.5mm hex wrench adjust the set screw clockwise to create more tension for the light head and counter clockwise to loosen. (Figure
I
)
2 of 4
3 of 4
Maintenance and Cleaning/Disinfection
Figure
WARNING: To avoid personal injury, ensure the dental light is off before proceeding with the following steps.
AVERTISSEMENT: Pour éviter les blessures, s’assurer que la lampe dentaire est éteinte avant d’effectuer les étapes suivantes.
Refer to Forest Operatory Equipment Asepsis Operator’s Guide for infection control procedures.
Input Voltage
16/18 VAC
Light Source
10 LEDs (1 W Each)
Spot Size
2.75in - 5.5in (70mm X 140mm)
Light Intensity, Lux (Adjustable)
8,000 - 35,000
Color Temperature
4,200 - 6,000K
Protection against electrical dangers
Class 1 device
, IPXO
Power Supply/Transformer
Continuous Duty
UL CERTIFIED
Product, SAFETY evaluated to ES60601-1
CAUTION: Federal law restricts this device to sale by or on the order of a dental practitioner licensed
by the law of the State in which he/she practices to use or order the use of the device.
MISE EN GARDE: Selon les lois fédérales, cet appareil ne peut être vendu que par un (ou à la demande
d’un) dentiste praticien autorisé par les législations de l’état dans lequel il ou elle exerce à utiliser ou
ordonner l’utilisation de cet appareil.
NOTE: Refer to Service Manual and Installation Instructions for complete technical descriptions.
REMARQUE: Consulter le manuel d’entretien et les instructions d’installation pour obtenir la
description technique complète.
Technical Description/Specifications
Indications for Use/Intended Use
The dental operatory light is a light source intended to
illuminate the oral cavity of dental patients for better visibility by a dental practitioner during diagnostic
and therapeutic treatment in a clinical environment. There are no contraindications for this product.
WARNING: Only the specified power supplies are to be used for safe operation of the dental light.
AVERTISSEMENT: Seules les alimentations spécifiées doivent être utilisées pour un fonctionnement
sécuritaire de la lampe dentaire.
Powertronix Power Supply AA-119298-ME (230/115 VAC, 244W, 50/60 Hz), AA119296-ME (230/115
VAC, 100W, 50/60 Hz) and Transformer AA-132400-ME (230/115 VAC, 100W, 50/60 Hz).
Refer to light installation instruction for locating power supply/transformer away from the vicinity of
water sources.
Refer to Forest General Information Operator’s Guide
for Glossary of Symbols, Terms, EMC/Electrical Safety
Declaration and User Guidance.
Accessories and Detachable/Replacement Parts
1. Power Supply/Transformer
2. Front Shield
3. Handles
WARNING: Do not stare into beam of light when light is on.
AVERTISSEMENT: Ne pas regarder directement le faisceau de lumière lorsque l’éclairage est en marche.
NOTE: Color temperature is subjective and varies between operators.
REMARQUE: La couleur de température est subjective et varie selon les utilisateurs.
Operating Light (continued)
Expected Service Life
With proper maintenance and service, Forest Dental products are
designed for a defined “service life” under normal use (based on approximately 50 patients per
week) of 5 years from the date of manufacture, with the exception of serviceable components. Some
components may become obsolete due to changes in technology or due to product improvements
and may necessitate product updates or upgrades. At the end of the defined service life, all products
require examination by a trained service technician prior to continued use. Following this, additional
examinations are required every 5 years.