Position light for oral cavity illumination using handles for focal length and direction.

• 

Turning On. Toggle switch (Figure 

A

) left. 

         = On           = Off

• 

Automatic On/Off. Lift light flex arm above and below 0˚.  

        (To deactivate contact your service technician.)

• 

Adjusting light intensity. Use knob (Figure 

B

) to adjust intensity. 

• 

Adjusting color temperature. Use knob (Figure 

C

)

 

to adjust color temperature.

Removing & Replacing Front Shield

 

CAUTION: Never touch the LED reflectors or lamps. If necessary, clean only with low pressure air.

 

MISE EN GARDE: Ne touchez jamais les réflecteurs ou les lumières LED. Si nécessaire, nettoyer uniquement avec de l’air à basse pression.

1.   Turn the 1/4 turn fasteners on front of shield clockwise. (Figure 

D

)

2.   Pull shield out with fasteners. See (Figure 

E

)

3.   Reverse to reinstall. See (Figure 

F

)

Removing & Sterilizing Handles

1.   Press in the locking tab. (Figure 

G

)

2.   To completely unhook, wiggle and pull handle outward while  

 

pressing in locking tab. 

3.   Autoclave @ 250° F (121° C) for 10 minutes or 273° F (134° C)  

 

for  4 minutes.

4.   Reverse steps 1 and 2 to reinstall.

Cleaning Front Shield & Reflector

17UJ

A

B

C

D

E

F

G

H

I

 

CAUTION: Abrasive or detergent materials such as alcohol, trichloroethylene, gasoline (petrol), turpentine or the like should not be used on any part of 

 

the light.

 

MISE EN GARDE: Les matériaux ou détergents abrasifs comme l’alcool, le trichloréthylène, l’essence, la térébenthine et toute autre substance semblable ne  

 

devraient pas être utilisés sur aucune partie de la lampe.

1.  Remove front shield. (Figures 

D & E

)

2.  Clean the front shield with a soft cloth moistened with mild soapy water. Using chemical disinfectants can damage components. The front

 

shield is not autoclaveable.

3.   If needed, clean reflector with low pressure air only.

LED Light

Operating Light

Light Friction Adjustment Screw

1.  Looking at the front of the light, remove the right side decorative trim from the back of the light by pulling it up with your fingers. This  

 

 

will reveal the friction adjustment screw. (Figure 

H

)

2.   Using a 1.5mm hex wrench adjust the set screw clockwise to create more tension for the light head and counter clockwise to loosen. (Figure 

I

)

2 of 4

3 of 4

Maintenance and Cleaning/Disinfection

Figure

 

WARNING: To avoid personal injury, ensure the dental light is off before proceeding with the following steps.

 

AVERTISSEMENT: Pour éviter les blessures, s’assurer que la lampe dentaire est éteinte avant d’effectuer les étapes suivantes.

Refer to Forest Operatory Equipment Asepsis Operator’s Guide for infection control procedures.

Input Voltage  

16/18 VAC

Light Source  

10 LEDs (1 W Each)

Spot Size  

2.75in - 5.5in (70mm X 140mm)

Light Intensity, Lux (Adjustable)  

8,000 - 35,000

Color Temperature  

4,200 - 6,000K

Protection against electrical dangers  

Class 1 device

, IPXO

Power Supply/Transformer  

Continuous Duty

     UL CERTIFIED  

Product, SAFETY evaluated to ES60601-1

CAUTION: Federal law restricts this device to sale by or on the order of a dental practitioner licensed 

by the law of the State in which he/she practices to use or order the use of the device.
MISE EN GARDE: Selon les lois fédérales, cet appareil ne peut être vendu que par un (ou à la demande 

d’un) dentiste praticien autorisé par les législations de l’état dans lequel il ou elle exerce à utiliser ou 

ordonner l’utilisation de cet appareil.
NOTE: Refer to Service Manual and Installation Instructions for complete technical descriptions.
REMARQUE: Consulter le manuel d’entretien et les instructions d’installation pour obtenir la  

description technique complète.

Technical Description/Specifications

Indications for Use/Intended Use 

The dental operatory light is a light source intended to 

illuminate the oral cavity of dental patients for better visibility by a dental practitioner during diagnostic 

 and therapeutic treatment in a clinical environment. There are no contraindications for this product.

         WARNING: Only the specified power supplies are to be used for safe operation of the dental light.

        

       AVERTISSEMENT: Seules les alimentations spécifiées doivent être utilisées pour un fonctionnement 

         sécuritaire de la lampe dentaire. 

 

        

Powertronix Power Supply AA-119298-ME (230/115 VAC, 244W, 50/60 Hz), AA119296-ME (230/115 

VAC, 100W, 50/60 Hz) and Transformer AA-132400-ME (230/115 VAC, 100W, 50/60 Hz).
Refer to light installation instruction for locating power supply/transformer away from the vicinity of 

water sources.

Refer to Forest General Information Operator’s Guide 

for Glossary of Symbols, Terms, EMC/Electrical Safety 

Declaration and User Guidance.

Accessories and Detachable/Replacement Parts

1.   Power Supply/Transformer

2.   Front Shield

3.   Handles

 

WARNING: Do not stare into beam of light when light is on.         

 

AVERTISSEMENT: Ne pas regarder directement le faisceau de lumière lorsque l’éclairage est en marche.

 

NOTE: Color temperature is subjective and varies between operators.   

 

REMARQUE: La couleur de température est subjective et varie selon les utilisateurs. 

Operating Light (continued)

Expected Service Life  

With proper maintenance and service, Forest Dental products are 

designed for a defined “service life” under normal use (based on approximately 50 patients per 

week) of 5 years from the date of manufacture, with the exception of serviceable components. Some 

components may become obsolete due to changes in technology or due to product improvements 

and may necessitate product updates or upgrades. At the end of the defined service life, all products 

require examination by a trained service technician prior to continued use. Following this, additional 

examinations are required every 5 years.

17UJ