1 of 4
Fusion Rear Pivot Mount
Installation Instruction
Tools Required: Standard and Metric Hex Key Set, Torque Wrench with Sockets, Wrench Set, Flat Head Screwdriver and Nitrile Gloves.
Alert Symbols
The safety notices associated with the following should be given special attention when they appear in maintenance,
operating and emergency procedures in the guide.
CAUTION indicates that a particular procedure or precaution must be followed to avoid system irregularity or possible damage to the product.
NOTE indicates special information to improve the ease of maintaining the product, or to clarify important information.
WARNING indicates that the personal safety of the patient, end user or technician could be compromised by disregarding the WARNING.
Not following instructions may result in an injury.
CAUTION: Field modifications that alter the electrical and/or mechanical safety of Forest products conflict with agency construction file
requirements and are not sanctioned by Forest.
CAUTION: To avoid electrostatic discharge (ESD) damage to electronics, always wear grounded wrist strap attached to the facility ground
(eg: Banana plug) when handling circuit boards, chair controls and cables or working in the area of the electrical circuits.
NOTE: Installation Verification: Refer to Operator’s Guide or Service Manual on our website at Forestdental.com; perform all operations.
If operations performing as intended installation complete. If operations not performing as intended, review steps in installation instructions to
confirm correct installation or call Forest Technical Support at 800-423-3555.
DTE Oregon, Inc.
Address 6200 ne cherry drive hillsboro oregon 97124
USA
Phone
800 . 423 . 3555
503 . 640 . 3012 |
Fax
503 . 693 . 9715 |
Digital
forestdental.com
Forest Dental Equipment
6.24.19
0097-026
Rev. -
4 of 4
Removing Chair Covers
6. Lift upper cantilever cover off the chair lift frame.
7. Disconnect indicator wire connector by pressing the release clip.
1. Position the chair’s back and base in the fully up position.
2. Disconnect chair from mains power.
NOTE: If covers are already removed proceed to rear pivot mount kit on page 2.
Reinstall Chair Covers
1. Lift up resting upper rear cover on chair
platform to reinstall lower cover. Using
both hands, spread the plastic cover and
push inward over bolts.
2. Reconnect the memory button wire
connector.
4. Using an 5/32” hex key, reinstall the screws
(2) on the sides of the upper rear cover.
3. Reinstall upper rear cover.
Wire connector
5. Turn rear side covers (2) clockwise to
reinstall.
Screws (2)
7. Reinstall pump cover using a 1/8” hex key.
8. Reconnect chair to mains power and verify operation.
6. Reinstall cantilever cover on the chair lift frame. Reconnect the indicator
wire connector (refer to step 2).
Wire connector
Release clip
1. Thread chair retaining bracket onto spiral
flex tubing.
2. Thread spiral flex tubing through utility
center retaining bracket.
3. Route tubing and wires through spiral flex
tubing into utility center.
Routing Tubing and Wires to Utility Center
Chair spiral
flex retaining
bracket
Tubings and
wires
Utility center
spiral flex
retaining
bracket
Wire connector
Release clip
3. Using a 1/8” hex key, remove
the screws (2) on the sides of
the pump cover.
4. Lift back corner and slide pump cover forward to disengage tabs
from brackets and remove.
Pump
cover
Screws (2)
5. Turn rear side covers (2) counter
clockwise to remove and set aside.
Tabs and
brackets(2)
Lift pump cover
CAUTION: A wire is connected to the indicator light. Be sure not to pull cover completely off before performing the next step.
Wire connector