Force AeroCollect User Manual Download Page 4

3.1 

Preparation of the sample chamber

The sample chambers are supplied in 5 ml protective tubes (see step 1 of 
the chamber insertion procedure below), which can easily be disinfected on 
the outside. To avoid contamination of the sample chambers, the protective 
tubes should not be opened before the user is in the location where the 
sample is to be collected. Be aware that the sample chambers are designed 
for single use as a disposable collection chamber.

A step by step guide for how to correctly insert the sample chamber in the 
device is given below:

1. 

 Open the protective tube and take out the sample 
chamber.  
 
The sample chambers are supplied in a protective 
tube, which can easily be disinfected on the 
outside. 

2. 

 Insert the sample chamber perpendicular to the 
top of the AeroCollect®, as shown in the picture.  
 
The arrows on the rear of the sample chamber 
indicate which side shall face up and which way 
the chamber shall be pushed in.

3. 

 Push the sample chamber all the way in. During 
this process it will move into position and be held 
by a spring. This ensures there is the necessary 
connection to the electrodes and the pump.

4. 

 AeroCollect® can collect a sample once the 
sample chamber is correctly inserted. 

Summary of Contents for AeroCollect

Page 1: ...User Manual...

Page 2: ...owever whenever the device is not being used it should be stored in the accompanying protective suitcase 2 Before you start First check that all of the parts are in the case see section 2 1 If some th...

Page 3: ...se contents 2 2 Overview of equipment Rear Front On Off Model Nr 000 Serie Nr 0000 FORCE Technology Venlighedsvej 4 2970 H rsholm WEEE CE mark Batteries 2 x AA Air out Sampling Green LED Wait Error Re...

Page 4: ...the sample chamber in the device is given below 1 Open the protective tube and take out the sample chamber The sample chambers are supplied in a protective tube which can easily be disinfected on the...

Page 5: ...it in the meantime The process should take less than 1 second 7 At 400 V the pump will start and the sampling procedure will be initiated which will be indicated by the green LED being lit while the r...

Page 6: ...ing Refer to the contact information at the end of this manual 4 2 Replacement of batteries AeroCollect is powered by two AA batteries The batteries are easy to replace Simply open the cover on the ba...

Page 7: ...cedure can be used to check if the sample chamber is blocked If the chamber is blocked the sound of the pump will not change either 4 5 Disinfection of AeroCollect Both the AeroCollect device and its...

Page 8: ...be too low and unable to power the pump If the batteries are drained of power and therefore cannot power the pump the pump will not start and the green LED will remain unlit and the red LED will remai...

Page 9: ...he device it must be sent in the accompanying case If the AeroCollect device stops working regardless of the reason immediately contact AeroCollect for the purpose of determining the cause of the malf...

Page 10: ...0 C Transport of collected samples temperature range 80 C to 30 C Normal use temperature range 20 C to 40 C 8 3 4 Recommended air humidity for sample chamber Transport and storage air humidity range 3...

Page 11: ...9 Contact information For more information about the use of AeroCollect contact FORCE Technology AeroCollect Venlighedsvej 4 2970 H rsholm Denmark Tel 45 43 25 14 00 aerocollect force dk...

Page 12: ...5098 3 EN FORCE Technology AeroCollect Venlighedsvej 4 2970 H rsholm Denmark Tel 45 43 25 14 00 aerocollect force dk aerocollect dk...

Reviews: