3.1
Preparation of the sample chamber
The sample chambers are supplied in 5 ml protective tubes (see step 1 of
the chamber insertion procedure below), which can easily be disinfected on
the outside. To avoid contamination of the sample chambers, the protective
tubes should not be opened before the user is in the location where the
sample is to be collected. Be aware that the sample chambers are designed
for single use as a disposable collection chamber.
A step by step guide for how to correctly insert the sample chamber in the
device is given below:
1.
Open the protective tube and take out the sample
chamber.
The sample chambers are supplied in a protective
tube, which can easily be disinfected on the
outside.
2.
Insert the sample chamber perpendicular to the
top of the AeroCollect®, as shown in the picture.
The arrows on the rear of the sample chamber
indicate which side shall face up and which way
the chamber shall be pushed in.
3.
Push the sample chamber all the way in. During
this process it will move into position and be held
by a spring. This ensures there is the necessary
connection to the electrodes and the pump.
4.
AeroCollect® can collect a sample once the
sample chamber is correctly inserted.