Summary of Contents for AeroCollect

Page 1: ...User Manual...

Page 2: ...owever whenever the device is not being used it should be stored in the accompanying protective suitcase 2 Before you start First check that all of the parts are in the case see section 2 1 If some th...

Page 3: ...se contents 2 2 Overview of equipment Rear Front On Off Model Nr 000 Serie Nr 0000 FORCE Technology Venlighedsvej 4 2970 H rsholm WEEE CE mark Batteries 2 x AA Air out Sampling Green LED Wait Error Re...

Page 4: ...the sample chamber in the device is given below 1 Open the protective tube and take out the sample chamber The sample chambers are supplied in a protective tube which can easily be disinfected on the...

Page 5: ...it in the meantime The process should take less than 1 second 7 At 400 V the pump will start and the sampling procedure will be initiated which will be indicated by the green LED being lit while the r...

Page 6: ...ing Refer to the contact information at the end of this manual 4 2 Replacement of batteries AeroCollect is powered by two AA batteries The batteries are easy to replace Simply open the cover on the ba...

Page 7: ...cedure can be used to check if the sample chamber is blocked If the chamber is blocked the sound of the pump will not change either 4 5 Disinfection of AeroCollect Both the AeroCollect device and its...

Page 8: ...be too low and unable to power the pump If the batteries are drained of power and therefore cannot power the pump the pump will not start and the green LED will remain unlit and the red LED will remai...

Page 9: ...he device it must be sent in the accompanying case If the AeroCollect device stops working regardless of the reason immediately contact AeroCollect for the purpose of determining the cause of the malf...

Page 10: ...0 C Transport of collected samples temperature range 80 C to 30 C Normal use temperature range 20 C to 40 C 8 3 4 Recommended air humidity for sample chamber Transport and storage air humidity range 3...

Page 11: ...9 Contact information For more information about the use of AeroCollect contact FORCE Technology AeroCollect Venlighedsvej 4 2970 H rsholm Denmark Tel 45 43 25 14 00 aerocollect force dk...