This device has been tested to meet the electrical and safety requirements of: 

IEC/EN 60601-1, IEC/EN 60601-1-2.
 
Reference to Standards: 

• EN 1060-1 /-3, NIBP-requirements
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC
• EN1060-4, NIBP clinical investigation
• AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements

 • 

 Batteries might leak chemicals if unused for a long time. Remove the 

batteries if you are not going to use the device for an extended period 

     (i.e., 3 months or more).

 • 

 Properly dispose of the batteries according to your local environmental 

regulations.

SYMBOL INFORMATION

DEALER’S  INFORMATION

 • 

 Storage condition:  -20°C to 70°C (-4°F to 158°F), between 10% to 95% relative 

humidity.

 • 

 Always store or transport the meter in its original storage case.

 • 

 Avoid dropping or heavy impact.

 • 

 Avoid direct sunlight and high humidity.

1. To clean the meter exterior, wipe it with a cloth moistened with tap water or a 

mild cleaning agent, then dry the device with a soft dry cloth. Do 

NOT

 flush 

with water.

2. Do 

NOT

 use organic solvents to clean the meter.

3. Do 

NOT

 wash the pressure cuff.

4. Do 

NOT

 iron the pressure cuff.

To avoid the meter attracting dirt, dust or other contaminants, wash and dry your 

hands thoroughly before use.

1. Connect AC adapter plug to DC adapter jack of the 

meter.

1. 

When the meter is off,

 remove AC adapter power 

plug from the electrical outlet.

2. Plug AC adapter power plug into an electrical outlet. 

Press 

ON/OFF

 to start the measurement.

2. Disconnect AC adapter plug from DC adapter jack of 

the monitor.

USING AC ADAPTER 

(only for TD-3124B)

CARING FOR YOUR METER

If you follow the recommended action but the problem persists, or error messages 

other than the ones below appear, please call your local customer service. Do not 

attempt to repair by yourself and never try to disassemble the monitor under any 

circumstances.

SYSTEM TROUBLESHOOTING

Connect AC Adapter to the Meter

Cleaning

Meter Storage

Remove AC Adapter from the Meter

Error Messages

MESSAGE                 

    

CAUSE 

      

     

   WHAT TO DO

Inflation or pressure error.

Please contact local customer 
service for help.

Refit cuff tightly and correctly. 
Relax and repeat the 
measurement. If error still 
remains, contact local 
customer service for help.

Replace the batteries.

Review the instructions and 
repeat the test. If the meter 
still does not work, please 
contact the local customer 
service for help.

Blood pressure 
measurement error.

Batteries are dead.

Problems with the meter.

POSSIBLE CAUSE

WHAT TO DO

Batteries exhausted.
Batteries incorrectly installed or 

absent. 

Replace the batteries.
Check that the batteries are correctly 

installed.

POSSIBLE CAUSE

WHAT TO DO

Movement during measurement.
Measurement taken just after 

exercise. 

Repeat measurement.
Rest at least 30 minutes before 

repeating measurement.

1. If no display appears after pressing       :

2. If the heart rate is higher/lower than user’s average:

Troubleshooting

3. If the result is higher/lower than user’s average measurement:

POSSIBLE CAUSE

WHAT TO DO

May not be in correct position while 

measuring.
Blood pressure naturally varies from 

time to time.

Adjust to the correct position to 

measure.
Keep in mind for next measurement.

4. If the cuff inflates again while measuring:

POSSIBLE CAUSE

WHAT TO DO

Cuff is not fastened.
If user’s blood pressure is higher than the pressure the device has inflated, the 

device will automatically increase the pressure and start to inflate again. Stay 

relaxed and wait for the measurement.

Fasten the cuff again.

Symbol

Referent

Clinical studies show that adult diabetes is often accompanied by elevated blood 

pressure. People with diabetes can reduce their heart risk by managing their 

blood pressure along with diabetes treatment*

  

.

Monitoring your routine blood pressure trend helps you to know your body 

condition. Human blood pressure naturally increases after reaching middle age. 

This symptom is a result of continuous ageing of the blood vessels. Further causes 

include obesity, lack of exercise and cholesterol (LDL) adhering to the blood 

vessels. Rising blood pressure accelerates hardening of the arteries, and the body 

becomes more susceptible to apoplexy and coronary infarction. 

DETAILED INFORMATION

Reference Values

SPECIFICATIONS

System performance

Blood pressure measurement performance

Four 1.5V AA alkaline batteries

130.5 mm (L) x 90 mm (W) x 55 mm (H), 

210 g without batteries.

24-43 cm (9.4-16.9 inches) 

with air tube 100 cm

Maximum 100 memory records (only for TD-3124B)

Automatic power off if system idle for 3 minutes 

5

o

C to 40

o

C (41

o

F to 104

o

F), between 15% - 93% RH

-25

o

C to 70

o

C (-13

o

F to 158

o

F), between 10% - 95% RH

four (4)1.5V AA size alkaline batteries (for TD-3124A/B)

DC +6V / 1A (max) via Power Plug (only for TD-3124B)

Power source:

Size of monitor w/o cuff:

Cuff Size:

Memory:

Power saving:

System operating conditions:

Meter storage conditions:

Power Supply Input:

0 - 300 mmHg

mmHg

60 mmHg -255 mmHg

30 mmHg -195 mmHg

40 -199 beats / minute

280 mmHg

±3 mmHg or ±2% of reading

±4% of reading

40 - 199 beat per minute  

Pressure range:

Measurement unit:

Systolic Measurement Range:

Diastolic Measurement Range:

Pulse Rate Measurement Range:

Maximum inflation pressure:

Accuracy of Pressure: 

Accuracy of Pulse rate:

Heart rate range:  

Consult instructions for use

Manufacturer

Serial number

Caution, consult accompanying documents

Type BF Equipment

Collection for electrical and electronic equipment

Distributed by

ForaCare Global Corp.

1F, No. 127, Wugong 2nd Rd., Wugu District,

New Taipei City 24888, Taiwan

Tel: +886-2-66253345 Fax: +886-2-66259878

www.foracare.com

Manufacturer

TaiDoc Technology Corporation

B1-7F., No. 127, Wugong 2nd Rd.,

Wugu Dist., New Taipei City 248, Taiwan(R.O.C.)

www.taidoc.com

1. We warrant this product to be free of defects in workmanship and materials

    within the said warranty period on the warranty certificate.

2. During the warranty period, if this product is found to be defective, you may bring it, 

    together with the purchase receipt and Warranty Certificate, on a carry-in basis to our 

    office during normal business hours for warranty service.  We will then repair or replace 

    defective parts or exchanging the whole product as we may choose, at no charge to 

    the original owner.  After such repair, replacement or exchange, the product will be 

    warranted for up to the remainder of the warranty period.

3.This warranty is valid only if the Warranty Certificate and Warranty Registration Card are 

   duly completed with date of purchase, serial number and dealer

s stamp, and if the 

   Warranty Registration Card is sent to our office not later than 14 days from the date of purchase.

4. This warranty is void if this product has been repaired or serviced by unauthorized person.  

    This warranty does not cover defects caused by misuse, abuse, accident, tampering, lack 

    of reasonable care, fire or any other acts beyond human control.

5. Except as stated in the above paragraphs, we disclaim all other warranties, implied or 

    expressed, including the warranties of merchantability or fitness for a particular purpose 

    with respect to the use of this product.  We shall not be liable for any direct, consequential 

    or incidental damages arising out of the use or inability to use this product.

4

The recommended blood pressure range is as below:

*2: National Heart, Lung, and Blood Institute, Diseases and Conditions 
*3: The Seventh Report of the Joint National Committee on Prevention, Detection, 

Evaluation, and Treatment of High Blood Pressure. NIH Publication. 2003. No. 03-5233 
*4: American Diabetes Association: The Diabetes-Heart Disease Link Surveying 

Attitudes, 

Classification

Systolic Pressure (mmHg) Diastolic Pressure (mmHg)

Hypotension*
Normal*
Pre-hypertension* 
Stage 1 Hypertension*
Stage 2 Hypertension*

Less than 90

Less than 120

120 – 139
140 – 159

160 or more

Less than 60
Less than 80

80 – 89
90 – 99

100 or more

2

3

3

3

3

SN  