This device has been tested to meet the electrical and safety requirements of:
IEC/EN 60601-1, IEC/EN 60601-1-2.
Reference to Standards:
• EN 1060-1 /-3, NIBP-requirements
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC
• EN1060-4, NIBP clinical investigation
• AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements
•
Batteries might leak chemicals if unused for a long time. Remove the
batteries if you are not going to use the device for an extended period
(i.e., 3 months or more).
•
Properly dispose of the batteries according to your local environmental
regulations.
SYMBOL INFORMATION
DEALER’S INFORMATION
•
Storage condition: -20°C to 70°C (-4°F to 158°F), between 10% to 95% relative
humidity.
•
Always store or transport the meter in its original storage case.
•
Avoid dropping or heavy impact.
•
Avoid direct sunlight and high humidity.
1. To clean the meter exterior, wipe it with a cloth moistened with tap water or a
mild cleaning agent, then dry the device with a soft dry cloth. Do
NOT
flush
with water.
2. Do
NOT
use organic solvents to clean the meter.
3. Do
NOT
wash the pressure cuff.
4. Do
NOT
iron the pressure cuff.
To avoid the meter attracting dirt, dust or other contaminants, wash and dry your
hands thoroughly before use.
1. Connect AC adapter plug to DC adapter jack of the
meter.
1.
When the meter is off,
remove AC adapter power
plug from the electrical outlet.
2. Plug AC adapter power plug into an electrical outlet.
Press
ON/OFF
to start the measurement.
2. Disconnect AC adapter plug from DC adapter jack of
the monitor.
USING AC ADAPTER
(only for TD-3124B)
CARING FOR YOUR METER
If you follow the recommended action but the problem persists, or error messages
other than the ones below appear, please call your local customer service. Do not
attempt to repair by yourself and never try to disassemble the monitor under any
circumstances.
SYSTEM TROUBLESHOOTING
Connect AC Adapter to the Meter
Cleaning
Meter Storage
Remove AC Adapter from the Meter
Error Messages
MESSAGE
CAUSE
WHAT TO DO
Inflation or pressure error.
Please contact local customer
service for help.
Refit cuff tightly and correctly.
Relax and repeat the
measurement. If error still
remains, contact local
customer service for help.
Replace the batteries.
Review the instructions and
repeat the test. If the meter
still does not work, please
contact the local customer
service for help.
Blood pressure
measurement error.
Batteries are dead.
Problems with the meter.
POSSIBLE CAUSE
WHAT TO DO
Batteries exhausted.
Batteries incorrectly installed or
absent.
Replace the batteries.
Check that the batteries are correctly
installed.
POSSIBLE CAUSE
WHAT TO DO
Movement during measurement.
Measurement taken just after
exercise.
Repeat measurement.
Rest at least 30 minutes before
repeating measurement.
1. If no display appears after pressing :
2. If the heart rate is higher/lower than user’s average:
Troubleshooting
3. If the result is higher/lower than user’s average measurement:
POSSIBLE CAUSE
WHAT TO DO
May not be in correct position while
measuring.
Blood pressure naturally varies from
time to time.
Adjust to the correct position to
measure.
Keep in mind for next measurement.
4. If the cuff inflates again while measuring:
POSSIBLE CAUSE
WHAT TO DO
Cuff is not fastened.
If user’s blood pressure is higher than the pressure the device has inflated, the
device will automatically increase the pressure and start to inflate again. Stay
relaxed and wait for the measurement.
Fasten the cuff again.
Symbol
Referent
Clinical studies show that adult diabetes is often accompanied by elevated blood
pressure. People with diabetes can reduce their heart risk by managing their
blood pressure along with diabetes treatment*
.
Monitoring your routine blood pressure trend helps you to know your body
condition. Human blood pressure naturally increases after reaching middle age.
This symptom is a result of continuous ageing of the blood vessels. Further causes
include obesity, lack of exercise and cholesterol (LDL) adhering to the blood
vessels. Rising blood pressure accelerates hardening of the arteries, and the body
becomes more susceptible to apoplexy and coronary infarction.
DETAILED INFORMATION
Reference Values
SPECIFICATIONS
System performance
Blood pressure measurement performance
Four 1.5V AA alkaline batteries
130.5 mm (L) x 90 mm (W) x 55 mm (H),
210 g without batteries.
24-43 cm (9.4-16.9 inches)
with air tube 100 cm
Maximum 100 memory records (only for TD-3124B)
Automatic power off if system idle for 3 minutes
5
o
C to 40
o
C (41
o
F to 104
o
F), between 15% - 93% RH
-25
o
C to 70
o
C (-13
o
F to 158
o
F), between 10% - 95% RH
four (4)1.5V AA size alkaline batteries (for TD-3124A/B)
DC +6V / 1A (max) via Power Plug (only for TD-3124B)
Power source:
Size of monitor w/o cuff:
Cuff Size:
Memory:
Power saving:
System operating conditions:
Meter storage conditions:
Power Supply Input:
0 - 300 mmHg
mmHg
60 mmHg -255 mmHg
30 mmHg -195 mmHg
40 -199 beats / minute
280 mmHg
±3 mmHg or ±2% of reading
±4% of reading
40 - 199 beat per minute
Pressure range:
Measurement unit:
Systolic Measurement Range:
Diastolic Measurement Range:
Pulse Rate Measurement Range:
Maximum inflation pressure:
Accuracy of Pressure:
Accuracy of Pulse rate:
Heart rate range:
Consult instructions for use
Manufacturer
Serial number
Caution, consult accompanying documents
Type BF Equipment
Collection for electrical and electronic equipment
Distributed by
ForaCare Global Corp.
1F, No. 127, Wugong 2nd Rd., Wugu District,
New Taipei City 24888, Taiwan
Tel: +886-2-66253345 Fax: +886-2-66259878
www.foracare.com
Manufacturer
TaiDoc Technology Corporation
B1-7F., No. 127, Wugong 2nd Rd.,
Wugu Dist., New Taipei City 248, Taiwan(R.O.C.)
www.taidoc.com
1. We warrant this product to be free of defects in workmanship and materials
within the said warranty period on the warranty certificate.
2. During the warranty period, if this product is found to be defective, you may bring it,
together with the purchase receipt and Warranty Certificate, on a carry-in basis to our
office during normal business hours for warranty service. We will then repair or replace
defective parts or exchanging the whole product as we may choose, at no charge to
the original owner. After such repair, replacement or exchange, the product will be
warranted for up to the remainder of the warranty period.
3.This warranty is valid only if the Warranty Certificate and Warranty Registration Card are
duly completed with date of purchase, serial number and dealer
’
s stamp, and if the
Warranty Registration Card is sent to our office not later than 14 days from the date of purchase.
4. This warranty is void if this product has been repaired or serviced by unauthorized person.
This warranty does not cover defects caused by misuse, abuse, accident, tampering, lack
of reasonable care, fire or any other acts beyond human control.
5. Except as stated in the above paragraphs, we disclaim all other warranties, implied or
expressed, including the warranties of merchantability or fitness for a particular purpose
with respect to the use of this product. We shall not be liable for any direct, consequential
or incidental damages arising out of the use or inability to use this product.
4
The recommended blood pressure range is as below:
*2: National Heart, Lung, and Blood Institute, Diseases and Conditions
*3: The Seventh Report of the Joint National Committee on Prevention, Detection,
Evaluation, and Treatment of High Blood Pressure. NIH Publication. 2003. No. 03-5233
*4: American Diabetes Association: The Diabetes-Heart Disease Link Surveying
Attitudes,
Classification
Systolic Pressure (mmHg) Diastolic Pressure (mmHg)
Hypotension*
Normal*
Pre-hypertension*
Stage 1 Hypertension*
Stage 2 Hypertension*
Less than 90
Less than 120
120 – 139
140 – 159
160 or more
Less than 60
Less than 80
80 – 89
90 – 99
100 or more
2
3
3
3
3
SN