![Fora TD-3261G Owner'S Manual Download Page 21](http://html1.mh-extra.com/html/fora/td-3261g/td-3261g_owners-manual_2316825021.webp)
21
QUALITY CONTROL TESTING
When should the control solution test be performed?
• if it is mandatory following the local regulations in your country,
• if you suspect the monitor or test strips are not working properly
• if your blood glucose test results are not consistent with how you
feel, or if you think the results are not accurate
• to practice testing process, or
• if you have dropped or think you may have damaged the monitor.
Performing a Control Solution Test
Step 1
Insert the test strip to turn on the monitor
Insert the test strip into the monitor. Wait
for the monitor to display the test strip and
blood drop symbol.
Step 2
Press
to mark this test as a control
solution test
With “QC” displayed, the monitor will store
your test result in memory. If you press
again, the “QC” will disappear and this test
is no longer a control solution test.
Summary of Contents for TD-3261G
Page 1: ...Owner s Manual FORA TD 3261G Blood Glucose Plus Blood Pressure Monitoring System...
Page 2: ...2...
Page 10: ...10 Monitor Overview Front Side Back Pressure Cuff 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18...
Page 20: ...20 Step 2 Press to switch between General AC PC and QC mode...