● Blood Pressure Measurement Performance
Pressure Range:
0 to 300 mmHg
Measurement Unit:
mmHg
Systolic Measurement Range:
60 to 250 mmHg
Diastolic Measurement Range:
30 to 180 mmHg
Pulse Rate Measurement Range:
30 to 199 beats / minute
Maximum Inflation Pressure:
280 mmHg
Accuracy of Pressure:
±3 mmHg or ±2% of reading
Accuracy of Pulse Rate:
±4% of reading
This device has been tested to meet the electrical and safety requirements of: IEC/
EN 60601-1, IEC/EN 60601-1-2, IEC/EN 61010-1, IEC/EN 61010-2-101, IEC/EN
61326-2-6.
Reference to Standards:
• IEC 60601-1 General requirement for safety
• IEC 60601-1-2 Requirements for EMC
• AAMI/ANSI /IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements