Operating Instructions  

EMGZ491.EIP 

Single-channel measuring amplifier for ETHERNET/IP 

EMGZ491.R.EIP for mounting on DIN rail 

EMGZ491.W.EIP for wall mounting     

 

Document Version 

2.5 

 03/2022 NS 

 Firmware Version 

V 2.0.4 

 EDS 

Datei 

FMS_TensionAmpilfier_EMGZ491_EIP_V2_1.eds 

 

 

Diese Bedienungsanleitung ist auch in Deutsch erhältlich. 

Bitte kontaktieren Sie Ihre lokale FMS Vertretung. 

 

© by FMS Force Measuring Systems AG, CH-8154 Oberglatt – All rights reserved. 

 

 

Summary of Contents for EMGZ491.EIP

Page 1: ...il EMGZ491 W EIP for wall mounting Document Version 2 5 03 2022 NS Firmware Version V 2 0 4 EDS Datei FMS_TensionAmpilfier_EMGZ491_EIP_V2_1 eds Diese Bedienungsanleitung ist auch in Deutsch erh ltlich...

Page 2: ...rnet Connections 10 5 CALIBRATION OF THE MEASURING SYSTEM 11 5 1 Offset Compensation 11 5 2 Calibration Adjusting the Gain Factor 11 5 3 Calibrating 12 5 4 Gain 12 5 5 Limit Value Violations 13 5 5 1...

Page 3: ...Operating Instruction EMGZ491 EIP 17 03 2022 3 10 DIMENSIONS 33 11 TECHNICAL DATA 34...

Page 4: ...ance data can endanger the safety and health of persons Work with respect to operation maintenance retrofit repair or setting the device described here must only be performed by expert personnel 2 1 P...

Page 5: ...e the mounting information on the following pages Observe the local installation regulations Improper handling of the electronics module can lead to damage to the sensitive electronics Do not work wit...

Page 6: ...rthermore device information parameters and system settings can be accessed via a web browser Offset compensation and system calibration can also be performed via the web browser 3 3 Scope of Delivery...

Page 7: ...Connect the force sensors to the amplifier see 4 5 Connect the amplifier to the supply voltage The voltage supply must be in the range of 18 to 36 VDC See 4 5 Perform offset compensation and calibrati...

Page 8: ...two force sensors can be connected to the EMGZ491 EIP When two sensors are used they are connected in parallel internally The force sensors and amplifier are connected using a 2x2x0 25 mm2 AWG 23 shi...

Page 9: ...nd the M12 plug must be loosened for board access You can slide out the board by approx 2 cm 1 in and loosen the terminal blocks for easier connection of the wires Figure 4 Pc board with removable ter...

Page 10: ...BA_Manual ai Warning Poor grounding can result in electric shocks for persons malfunctions of the overall system or damage to the measuring amplifier Proper grounding must always be ensured Note Cable...

Page 11: ...rocedure 5 2 Calibration Adjusting the Gain Factor Calibration is used for matching the gain factor with the force sensors After calibration the displayed force corresponds to the force effectively af...

Page 12: ...ler the applied force is not relayed to the sensors 1 1 Thus the measured force does not correspond to the effectively applied force To correct for this error the measured force is amplified using a f...

Page 13: ...hen calibrating via web browser 5 5 Limit Value Violations The amplifier checks the analog input and output for limit value violations At the input it is checked using the input voltage whether the fo...

Page 14: ...small over and or underflow Thus the overflow triggers when the theoretically calculated output value of 10 05 V is reached For underflow the value is 0 05 V When these limit values are reached the re...

Page 15: ...a quick start guide are available in the download section on www fms technology com The Quick Start Guide describes how to integrate the EMGZ491 492 into the PLC development environment EMGZ49x with...

Page 16: ...Configuration The EMGZ491 EIP can be configured either via the web interface or via ETHERNET IP The parameters Low pass filter active and Low pass filter analog output active cannot be accessed via th...

Page 17: ...Operating Instruction EMGZ491 EIP 17 03 2022 17...

Page 18: ...on Unit Here you select which unit of measurement is used The label located on the sensor will indicate the nominal force in Newtons Note This input will also affect the unit of the cyclic process dat...

Page 19: ...ing force capacity is installed in the measuring roller E g if two 500 N force sensors are installed in the roller 1 000 N must be entered In the case of a one sided measurement hence is one 500 N for...

Page 20: ...ow pass filter for the analog output is indicated This parameter cannot be accessed via the web interface Min 0 Max 1 Specified value 1 0 no inactive 1 yes active Limit frequency low pass filter analo...

Page 21: ...to 16383 Value format The value is interpreted as integer without decimal place E g 1000 1000 ADC raw value Actual value in Newton Filtered actual value converted into Newton Data type long signed 32...

Page 22: ...decimal places E g unit set to kN 100000 100 000 kN 100 kN Value format for g The value is interpreted as decimal number with 1 decimal place E g unit set to g 12340 1234 0g 1234g Status The status co...

Page 23: ...re transmitted using acyclic data traffic Parameter ID Description 10241 Unit Access type R W Parameter command unit Data type byte unsigned 8 bit Value range 0 to 4 0 N 1 kN 2 lb 3 g 4 kg Value forma...

Page 24: ...adjusted value is lost on a restart This filter is switched on after a restart This parameter cannot be accessed via the web interface Access type R W Parameter command low pass filter actual value ac...

Page 25: ...ritten to the device A reset is not necessary Every time a 1 is written the offset is evaluated 10248 Calibration Calibrates the amplifier to the weight in Newton which is handed over here It must mat...

Page 26: ...d value is lost on a restart This filter is switched on after a restart This parameter cannot be accessed via the web interface Access type R W Parameter command low pass filter analog output active D...

Page 27: ...e a Connect took place Figure 8 Read Write cycle EMGZ491_EIP_Grafik ai A computer with the respective application can now request a read or write on the data model of the controller The controller exe...

Page 28: ...into the web browser the homepage appears Figure 9 Homepage with device information The homepage provides information about general device properties such as serial number and software version The men...

Page 29: ...GZ491 EIP 17 03 2022 29 9 2 Parameter Settings The Parameters page offers the option to configure the amplifier via the web interface In the ETHERNET IP environment this is usually done from the PLC F...

Page 30: ...ing this page the offset can be adjusted and then the calibration executed These functions are also available via PLC If the values for offset and gain are known they can be directly assigned to the r...

Page 31: ...17 03 2022 31 9 4 Ethernet Settings This page shows the current TCP IP configuration It cannot by changed via the web interface but read only Changes to this configuration can only be applied via the...

Page 32: ...03 2022 32 9 5 System Settings Using the System Settings page the firmware version can be seen and new firmware can be loaded as well Figure 14 System settings You can find the current firmware files...

Page 33: ...ting Instruction EMGZ491 EIP 17 03 2022 33 10 Dimensions Figure 15 EMGZ491 R EIP housing for DIN rail mounting EMGZ491_EIP_Grafik ai Figure 16 EMGZ491 W EIP housing for wall mounting EMGZ491_EIP_Grafi...

Page 34: ...olution 32 768 digit 16 bit D A converter resolution 0 to 4 096 12 bit Measuring inaccuracy 0 05 FS Connector for interface EMGZ 491 R 2 x RJ 45 EMGZ 491 W 2 x M 12 4 pole D coded Configuration via ET...

Page 35: ...nd actual value in config unit Status Acyclic Services Get_Attribute_Single Set_Attribute_Single according to CIP spezification Volume 1 and Volume 2 Ring Topology DLR Device Level Ring Predefined Sta...

Page 36: ...s technology com www fms technology com FMS USA Inc 2155 Stonington Avenue Suite 119 Hoffman Estates IL 60169 USA Tel 1 847 519 4400 Fax 1 847 519 4401 fmsusa fms technology com FMS UK Aspstrasse 6 81...