Operating manual EMGZ300
10
7
Setting into operation
7.1
View of the setting elements
fig. 5: Setting elements for all channels
E300005e
7.2
Configuring gain and offset
Gain and offset are configured by jumpers. The settings can be made individually for each
channel.
Gain and offset variable
If gain and offset will be adjusted on the
electronic unit, the jumpers will be set to
„variable“ (fig. 6). Gain and offset then may
be adjusted using the trimmers (refer to „7.5
Calibration of the measuring amplifier“).
Gain and offset fixed
If the signal is processed by a control unit (for
ex. a PLC), configuration with fixed gain and
offset is recommended (offset = 0; gain =
1111). Therefore, the jumpers will be set to
„fixed“ (fig. 7). This configuration takes the
highest advantage of the good thermal
characteristics of the hybrid module. The
output signal will be 10V resp. 20mA when
nominal force of the sensor is applied.
fig. 6: Jumper setting for variable gain
and offset
E300006a
fig. 7: Jumper setting for fixed gain
and offset
E300006b