2.3 Step Motors:

FMI STH and STQ standard pumps are supplied 
with a 23 frame, 1.8°/ step, hybrid 8  lead step motor.  
For  wiring diagram and detailed motor specifications    
see  

Figure F

2.4 Rotational Sensor:

FMI STH and 

STQ standard pumps are supplied with an LED 
rotational sensor.  For wiring diagram and sensor 
details see 

Figure G . 

2.4.1

For sensor use with FMI SCST-01 Step 

Motor Controller 
see SCST-01 instructions.

3.0

CALIBRATION AND TESTING

3.1

Flush the system: clean with alcohol 

or other suitable cleaning/wetting agent 
before using the hydraulics and pump for the 
first  time.  Prime pump with fluids to assure 
system is free of air bubbles.

3.2 

Place pump in the standard "home" 

position with the piston-flat perpendicular to 
the inlet port.The inlet port will be on the left 
when the motor rotates clockwise (see  

Figure B 

for the STH pump and  

Figure C 

for 

the STQ pump).

OUTLET

STQ PISTON FLAT - SENSOR / FLAG

RELATIONSHIP TO INLET AND OUTLET PORTS

CLOCKWISE
MOTOR
ROTATION

PISTON
FLAT

PISTON
FLAT

INLET

CLOCKWISE
MOTOR
ROTATION

OUTLET

ADJUSTMENT

SCREW

FLAG

SENSOR

STH PISTON FLAT - SENSOR / FLAG

RELATIONSHIP TO INLET AND OUTLET PORTS

INLET

3.3 Check the pump's dispense volume

by 

cycling the pump for one complete 360°
revolution (one complete rotation). Dispense onto 
an analytical pan balance and observe the volume.  

See Figure D. 

Readjust as necessary.

REMARK

: When dispensing water-like solutions

assume 1 ml = 1 gram.  A correction factor will 
be required for other fluids.

PAN BALA

INLET

OUTLET

WATER SUPPLY

CLOCKWISE

ROTATION

Figure B

Figure C

Figure D

pan Balance

2

5 Aerial Way, Suite 500, Syosset NY 11791 Tel: 516-922-6050 Fax: 516-624-8261 

Website: www.fmipump.com   Email: [email protected]

ST431-02

Summary of Contents for STH00

Page 1: ...s or liability for any misuse that will cause damage to product and or injury to the individual Power cables and leads should not be bent pulled or inserted by excessive force Otherwise there is a threat of electrical shock or fire Replace any inline fuses only with fuse rating as specified by Fluid Metering Inc When pump drive is under operation never point discharge tubing into face or touch any...

Page 2: ...inlet port will be on the left when the motor rotates clockwise see Figure B for the STH pump and Figure C for the STQ pump OUTLET STQ PISTON FLAT SENSOR FLAG RELATIONSHIP TO INLET AND OUTLET PORTS CLOCKWISE MOTOR ROTATION PISTON FLAT PISTON FLAT INLET CLOCKWISE MOTOR ROTATION OUTLET ADJUSTMENT SCREW FLAG SENSOR STH PISTON FLAT SENSOR FLAG RELATIONSHIP TO INLET AND OUTLET PORTS INLET 3 3 Check the...

Page 3: ...ion 3 Splashing can usually be avoided by modifying dispense tip to larger I D and or decreasing dispense speed 4 Hanging drop at dispense tip can be avoided by a Use of rigid discharge tubing b Small dispense tip c Increasing speed 5 High viscosity dispensing requires a Large suction tubing b Pressurizing suction reservoir 6 Bubbles in discharge a Suction fitting leak b Cavitation 7 Need Help Cal...

Page 4: ...SERIES BIPOLAR 4 LEAD LEAD COLOR CONNECTION DRIVER PIN CONNECTION 1 8 DEGREES OF ROTATION 2 1 1 3 2 3 2 2 5 5 DEGREES C PER WATT 0 0017 0 012 OZ IN SEC KG M 10 4 0 0 028 OZ_IN Nm OZ_IN Nm 89 0 63 63 0 44 82 0 58 2 3 1 06 0 206 2 5 89 0 63 116 0 82 2 3 0 53 0 206 3 5 89 0 63 116 0 82 9 2 2 12 0 412 1 75 OZ_IN Nm OZ_IN Nm mH OHMS Vpk RAD SEC AMPS UNIPOLAR 6 LEAD PARALLEL SERIES UNITS FMI STEP MOTOR ...

Reviews: