Section IV – Factor Selection for Preliminary Proving
MN06146 Issue/Rev. 0.1 (7/10)
7
Remember the Rules
1.
The maximum input frequency accepted by Ac-
cuLoad II directly from the meter is 3,000 Hz.
The maximum input frequency accepted by Ac-
cuLoad III directly from the meter is 10,000 Hz.
2.
The range of allowable Input Resolution Settings
is 0 to 9,999 pulses per whole unit of display
(i.e., gallons, liters, etc.) on the AccuLoad II, and
0 to 99999.999 on the AccuLoad III.
3.
The range of allowable Meter Factor Settings is
0.0001 to 9.9999 for the AccuLoad II and
0.00001 to 9.99999 for the AccuLoad III. Realis-
tic Meter Factors would be between 0.85 and
1.25.
Reproving the Meter
If reproving the system, use the factors already set
into the AccuLoad and proceed to the next section,
Factor Optimizing.