8.0  Application with low temperature

 

 

The VITRIOX GAS device not only allows the setting of different time sequences in the 
individual applications, but also different temperature profiles. 
 
These two parameters enable a time‐optimized and/or material‐specific adjustment of the 
fusion process. The device thus allows a largely individual adaptation to different 
requirements. 
 
In the preheating step, e.g. a low temperature can be set to e.g. to dry the crucible contents, 
or to perform an oxidation. Thereafter (from prefusion on), the normal melting temperature 
is used. 
 

9.0  Using the platinum ware 

 

Place the platinum crucible with sample into the receptacle holes in the crucible holder.   

 

Place the moulds onto the ceramic bars and push them up to the ceramic stop. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Hang the crucibles on the 
notches and push the crucible to 
the left until it is in the correct 
position. 

Summary of Contents for VITRIOX GAS 2XRF

Page 1: ...XANA GmbH Co KG Borschelstr 3 47551 Bedburg Hau Germany Tel 49 2821 48011 10 Fax 49 2821 48011 99 info fluxana de www fluxana com Service and Support FLUXANA GmbH Co KG Borschelstr 3 47551 Bedburg Hau...

Page 2: ...unction key Load 5 1 3 Function key Modify 5 1 4 Function key Settings 6 0 Copying an Application 7 0 Changing the fusion times during the running application 8 0 Application with low temperature 9 0...

Page 3: ...mbar Minimum maximum gas pressure 200 500 mbar 200 500 mbar 200 500 mbar Recommended oxygen pressure 2 bar 2 bar 2 bar Maximum oxygen pressure 2 bar 2 bar 2 bar Maximum air consumption 1 m3 h 2 m3 h 3...

Page 4: ...ar view 1 Gas connection hose fitting 9 mm outer diameter 2 Air pressure gauge 3 Air connection hose fitting 9 mm outer diameter 4 Oxygen connection hose fitting 6 mm outer diameter 5 Power connection...

Page 5: ...FXBA 0096 10 5 18 2 2 Front view 1 Touch display 2 Emergency stop button 3 Machine power switch Main switch Power 4 USB port 1 2 3 4...

Page 6: ...FXBA 0096 10 6 18 2 3 Top view 4 Stations 1 Crucible burner 2 Mold burner 3 Crucible 4 Mould 5 Ceramic bar for mould 6 Ceramic stop for mould 2 1 3 4 5 6...

Page 7: ...ressed air to the Air hose fitting with a hose clamp Inner diameter of the hose for compressed air is 9 mm Air pressure min 4 bar max 8 bar Connect the oxygen to the Oxygen hose fitting with a hose cl...

Page 8: ...y the pilot burner After pouring the melt out of the crucible into the moulds they are cooled first with low pressure air cooling 1 and then cooled in the next two steps with increasing pressure The f...

Page 9: ...d after pressing the Power button 5 1 Function Keys Appl Application can be selected Station stations can be switched on and off Modify Applications can be changed Load selected application is read St...

Page 10: ...automatically loaded when the device is switched on After selecting the application type application number 0 9 and confirm with ENTER read application with Load 5 1 3 Function key Modify With Modify...

Page 11: ...e new window select Admin enter the password and confirm with Enter Press Save to store all new entries these are retained even after the device has been switched off Exit this page with Menu 5 1 4 Fu...

Page 12: ...um temperature before pouring Example for shortening the Mainfusion time The device is currently in the Mainfusion 1 Target 240 sec After inspection the main fusion should end after 180 seconds Press...

Page 13: ...gely individual adaptation to different requirements In the preheating step e g a low temperature can be set to e g to dry the crucible contents or to perform an oxidation Thereafter from prefusion on...

Page 14: ...ds Mainfusion 2 240 seconds Agitating off 10 seconds Agitating on 10 seconds Pouring 7 seconds Cooling complete 360 seconds 60 240 60 Application 1 Preheating 240 seconds Prefusion 540 seconds Mainfus...

Page 15: ...ntric upwards out of the motor plate Put round belt on the motor plate and put the eccentric back in the holes of the motor plate Now place the round belt over the grooves of the drive pulleys Figure...

Page 16: ...y to find out if the toothed belt is in the right position If not adjust until everything fits Now tighten the screws slightly and insert the eccentric into the motor plate If the shafts rotate easily...

Page 17: ...in disregard of instructions furnished by the producer In no event shall damages for defective goods exceed the purchase price of the goods and the producer shall not be liable for incidental or cons...

Page 18: ...directives by FLUXANA GmbH Co KG Borschelstr 3 47551 Bedburg Hau This declaration relates exclusively to the machinery in the state in which was placed on the market and excludes components which are...

Reviews: