6.3  Setting the calibrating factor 

After  pressing the "Mode" key,  the  calibrating factor set  appears  in the display.  Pressing the 
"Total" key results in the calibrating factor being counted up in 0.010 steps and pressing the 
„Reset“ key in it being counted down accordingly. 
Press the "Mode" key and the set calibrating factor is taken on.  

6.4  Resetting all settings (initialisation) 

The effect of pressing and holding the "Total" and "Reset" keys at the same time when setting 
the  measurement  unit  or  calibrating  factor  is  to  delete  all  the  meter  values  and  carry  out  an 
initialisation.  
 
Following values are initialised: 

Measurement unit:   

Litre 

Calibrating factor:   

1.000 

Delivery quantity:    

0.00 litres 

Totaliser:  

 

 

0 litres 

6.5  Ending the programming mode 

In order to end the programming mode, the "Mode" key is again to be pressed after setting the 
calibrating factor. The meter then reverts to the basic state. When no key is pressed for over  
5 minutes in the programming mode, the meter automatically returns to the basic state. 
 

 

Summary of Contents for FMT 50 PP

Page 1: ...roduct description 2 2 Safety instructions 2 3 Technical Description 4 4 Operation 4 5 Fault monitoring 5 6 Meter programming 5 7 Spare parts construction and spare parts list 6 8 Dimensional Drawing...

Page 2: ...ir which in the case of volume flow transfers the meter pulses to a reed switch on the evaluation electronic system 2 Safety instructions The device is a state of the art piece of equipment and has be...

Page 3: ...erator If the personnel do not have the required knowledge they must be trained and instructed The operator must also ensure that the contents of the operating instructions are properly understood by...

Page 4: ...e LC display indicates the metered volume since the last reset The display has three digits before the decimal point and two after 0 01 litres is the minimum digital step The litre measurement unit op...

Page 5: ...ctor can be established a delivery must be made into a satisfactorily accurate vessel or undertaken by means of a reference meter The new calibrating factor is calculated as follows shown delivered ex...

Page 6: ...olding the Total and Reset keys at the same time when setting the measurement unit or calibrating factor is to delete all the meter values and carry out an initialisation Following values are initiali...

Page 7: ...uction and spare parts list Spare parts list Pos Description PP 10 Cover complete Battery A05 00 001 20 Measuring chamber typeplate A05 00 002 30 Battery 940 04 033 40 Four cylinder head screws DIN 91...

Page 8: ...range 0 8 40 mPa s Operating pressure 4 bar Burst pressure 9 bar Accuracy uncalibrated 2 Accuracy calibrated 1 Repeat accuracy 0 5 Nominal width 1 External thread Protection category IP 65 Battery Li...

Page 9: ...a construction du Compteur de D bit FLUX FMT 50 de mat riaux et versions diff rents est conforme aux dispositions r glementaires suivantes 1 EG Richtlinie Maschinen EC Machinery Directive Directive CE...

Page 10: ...10 Talweg 12 75433 Maulbronn Telefon 07043 101 0 Telefax 07043 101 444 Fax International 49 7043 101 555 info flux pumpen de www flux pumpen de...

Reviews: