Visual IR Thermometer
Main Menu
13
ROIMeasure
Use this setting to determine the measure area to help with temperature measurement. In the
measure area, high and low temperature markers automatically track the highest temperature and
lowest temperature. The images in
show the types of measure area that you can select.
Figure 2. Types of ROIMeasure Areas
Alarm
Use this setting to adjust the alarm parameters.
shows the options in the Alarm submenu.
To set the alarm:
1. In the Alarm submenu, push
O
to turn on or turn off the alarm.
2. When the alarm is on:
a. Push
X
to highlight
High
.
b. Push
O
to turn on the high-temperature alarm.
c. Push
X
to highlight
High temperature value
.
d. Push
O
to begin editing and push
W
or
X
to adjust the high temperature value.
e. Push
O
to stop editing and save the changes.
Table 7. Alarm Menu
Option
Suboption
Description
OFF
Turn off the alarm.
ON
High
Turn on or turn off the high-temperature alarm; set
the high-temperature alarm value.
High temperature value
Low
Turn on or turn off the low-temperature alarm; set the
low-temperature alarm value
.
Low temperature value
Center
Center with full screen
Center with small
spot box
Center with moderate
spot box