Model 90100
Operators Manual
4-4
Low Battery Indication at the Display
Figure 4-2
Low Battery Flowchart
Yes
No
Yes
Yes
No
Yes
No
No
35360A TRACKER Display is faulty and
will require repair or replacement.
Replace 35360A TRACKER
Fuses OK?
Check AC line fuses at rear panel
of 35360A TRACKER Display.
AC line voltage present?
Verify voltage present at AC line
receptical by connecting and
powering some other device.
Commence operation from AC line
power or charge batteries.
Low battery
annuciator displayed?
Connect 35360A TRACKER
Display to an AC power source
and press the POWER key.
Low Battery annuciator is displayed on
35360A TRACKER Display's VFD.
Operating from
AC power source?
When powering the 35360A TRACKER Display from
an AC line, always use a 3-wire grounding type line cord
h
as the one supplied with the 90100 TRACKER System.
In addition, only connect the 35360A TRACKER Display to
AC power outlet employing a third wire safety
d
Always turn the 35360A TRACKER Display's power off
prior to connecting or disconnecting the interface
bl
Failure to do so may result in an electrical
h
k
The Model 35360A TRACKER Display and 35300A
Detector contain no internal operator serviceable
Disassembly of either unit could result in electrical
h
k
WARNING
Display's AC line fuses.
Refer to Section 5.3.