Fluke T5-1000 Instruction Sheet Download Page 3

Battery Replacement

Observe the polarity markings inside the battery compartment.

+

+

Do not 
overtighten screw

Disconnect probes 
from Voltage

Switch OFF

2 AA

+

+

ex4.eps

Replacement Parts

Test Lead Assembly.
Replace only with Fluke double-insulated leads (

T

)

See "Accessories" for available probes.

PN 648029

Battery Door

PN 1626588

Battery Door Screw

PN 1618578

Accessories

H5 Belt Holster
TP1 Probe Set, Flat-Blade
TP4 Probe Set, 4 mm round
Service Information Sheet, PN 686953

To contact Fluke, call:

USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe:  +31 402-678-200
Japan:  +81-3-3434-0181
Singapore:  +65-738-5655
Anywhere in the world:  +1-425-446-5500

Visit Fluke’s Web site at www.fluke.com.

 

 

Specifications

Specifications subject to change without notice.

Calibration:  One-year calibration cycle.

Maximum Voltage Between any Terminal and Earth Ground:
T5-600:  600 V rms, Overvoltage  Category III; Pollution Degree 2
T5-1000:  1000 V rms, Overvoltage Category III, also 600 V rms
overvoltage Category IV, Pollution Degree 2

Maximum Voltage between Current Fork and Earth Ground: T5-600
and T5-1000: 1000 V rms, Overvoltage Category III, also 600 V rms,
Overvoltage Category IV, Pollution Degree 2, (This dual rating applies
to the current fork only.)

Temperature:  Operating: -10 

°

C to +50 

°

C (14 

°

F to 122 

°

F);

Storage: -30 

°

C to +60 

°

C (-22 

°

F to +140 

°

F)

Altitude:  Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)

Relative Humidity:  0 % to 95 %, 5 

°

C to 30 

°

C (41 

°

F to 86 

°

F); 0 % to

75 %, 30 

°

C to 40 

°

C (86 

°

F to 104 

°

F); 0 % to 45 %, 40 

°

C to 50 

°

C

(104 

°

F to 122 

°

F)

Battery Type and Life:  AA (2); 360 hours continuous with alkaline;
125 hours continuous with zinc chloride

Shock, Vibration:  1 m drop at -10 

°

C to 50 

°

C (14 

°

F to 122 

°

F) per

ANSI/ISA-S82.01-1994 and EN 61010-1: 1995. Random vibration per
MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g
maximum).

Surge Protection:  T5-600 6 kV per IEC 61010;

T5-1000 8 kV per IEC 61010

Enclosure Rating:  IP 52 per IEC 60529, no vacuum applied

RF Field Specification:  0.5 % full scale + (specified accuracy) at
3 V / m

Safety:  Complies with ANSI/ISA-S82.01-94, UL classified to IEC
61010, CSA/CAN C22.2 No.1010.1-92, and EN 61010-1: 1995.

EMC:  EN 61326

Certifications:

In accordance with
IEC 61010-1 54CJ

Summary of Contents for T5-1000

Page 1: ...ersonnel Use caution when working above 30 V ac rms 42 V ac peak or 60 V dc When using the probes keep your fingers behind the finger guards on the probes Connect the common test lead before you connect the live test lead Disconnect the live test lead first The display will not display hazardous voltages in display hold The voltage indicator LED continues to operate Symbols B AC alternating curren...

Page 2: ...display hazardous voltages in display hold The voltage indicator LED continues to operate To exit the display hold mode press and release HOLD again or turn the rotary switch to a new position Automatic Power Down The tester turns off automatically if you do not turn the rotary switch or press HOLD for approximately 45 minutes To resume operation turn the tester off then turn the tester on Low Bat...

Page 3: ...voltage Category IV Pollution Degree 2 Maximum Voltage between Current Fork and Earth Ground T5 600 and T5 1000 1000 V rms Overvoltage Category III also 600 V rms Overvoltage Category IV Pollution Degree 2 This dual rating applies to the current fork only Temperature Operating 10 C to 50 C 14 F to 122 F Storage 30 C to 60 C 22 F to 140 F Altitude Operating 2000 m 6562 ft Storage 10 000 m 32808 ft ...

Page 4: ...and lighting systems in large buildings CAT IV protects against transients from a primary supply such as an electricity meter or an overhead or underground utility service Limited Warranty Limitation Of Liability This Fluke product will be free from defects in material and workmanship for two years from the date of purchase This warranty does not cover fuses disposable batteries or damage from acc...

Reviews: