Multifunction PV Analyzer
Specifications
5
AC/DC Current with i100 Clamp
i100 Clamp Tolerances
AC/DC Power Measurement (with i100 Clamp)
Safety
SMFT-1000 ...................................................... IEC 61010-1 Pollution Degree 2
IEC 61010-2-034 CAT III 1000 V dc, CAT III 700 V ac
i100 Current Clamp....................................... IEC 61010-2-032, Type D (for insulated conductors), 1000 V
Accessories..................................................... IEC 61010-031
TL1000-MC4................................................ CAT III 1500 V, 20 A
TP1000 Remote Probe
with cap....................................................... CAT IV 600 V, CAT III 1000 V, 10 A
without cap ................................................ CAT II 1000 V, 10 A
TL1000 Test Leads..................................... CAT III 1000 V, 10 A
TL1000/30M Test Lead............................. CAT III 1000 V, CAT IV 600 V,
5 A (on reel) 10 A (fully extended)
TP74 Test Probes
with cap....................................................... CAT IV 600 V, CAT III 1000 V, 10 A
without cap ................................................ CAT II 1000 V, 10 A
AC285 Alligator Clips ................................ CAT III 1000 V, 10 A
Display Range
Measurement Range
Resolution
Accuracy
(DC, AC 50 Hz/60 Hz)
0.0 A dc to 100 A dc
1.0 A dc to 100 A dc
0.1 A
± (5 % + 2 Digit)
[1]
0.0 A ac to 100 A ac TRMS
1.0 A ac to 100 A ac TRMS
[1] i100 clamp tolerance not inclusive. See
Measurement Range
Output Signal
Accuracy
(DC, AC 50 Hz/60 Hz)
Maximum Hysteresis
1 A to 100 dc or ac <1 kHz
10 mV/A ac/dc
± (1.5 % + 0.5 A)
±0.4 A
Display Range
Measurement Range
Resolution
Accuracy
(DC, AC 50 Hz/60 Hz)
0.0 V ac to 700 V ac
5.0 V ac to 700 V ac
0.1 V
± (2.5 % + 2 Digit)
0.0 V dc to 1000 V dc
5.0 V dc to 1000 V dc
0.0 A ac/dc to 100 A ac/dc 1.0 A ac/dc to 100 A ac/dc 0.1 A
± (5 % + 6 Digit)
0 W/VA to 100 kW/kVA
5 W/VA to 100 kW/kVA
1 W / VA; 1 kW / kVA
± (7.5 % VI + 0.6 V + 0.2 I)