![Fluke QED 6 User Manual Download Page 35](http://html1.mh-extra.com/html/fluke/qed-6/qed-6_user-manual_2310046035.webp)
Operation
Adjusting
Display
Contrast
2
2-9
Adjusting Display Contrast
Display contrast on the Analyzer may be adjusted to optimize viewing of
menus and test data.
To set the display contrast:
1.
From
Main Menu 2
, press the
UTIL
softkey to display the following:
fcf055.eps
2.
Press the
DISP
softkey to access the display
Contrast
menu:
fcf057.eps
3.
Press the
+
softkey to increase the numerical value and decrease the
contrast; press the
–
softkey to decrease the numerical value and increase
the contrast. The default is 5.
4.
Press the
esc
softkey to store the last displayed value in memory.
Measuring Defibrillator Energy
To measure defibrillator energy:
1.
Power up the defibrillator to be tested and select the energy output
according to the manufacturer’s instructions.
2.
Power up the Analyzer.
3.
From
Main Menu 1
, press the
ENRG
softkey to access the
Energy