Introduction 

The PV350 Pressure/Vacuum Transducer Module (the module) 

measures gauge pressure; the difference between atmospheric 

(barometric) pressure and the pressure (or vacuum) applied to the 

pressure port on the transducer. It then converts that measurement to 

1 mV dc per unit. A module and a high-performance digital multimeter 

(DMM) with min/max function become a datalogger capable of measuring 

peak high and peak low.  

Box Contents 

Pressure/Vacuum Transducer Module

1/4 in. male flare to 1/4 in. female pipe thread connector

9 V alkaline battery

Instruction Sheet

Quick Reference Guide

Safety Information 

Warning 

To avoid eye injury, always wear approved safety glasses 

when working with this instrument.  

To avoid injury from pressure explosion: 

Be sure that pressure is removed before

disconnecting any hoses or tubing from the PV350 

transducer. 

Follow accepted safety procedures for systems on

which pressure or vacuum measurements are made. 

Make sure the PV350 is serviced by a qualified

technician. 

Summary of Contents for PV350

Page 1: ...ith min max function become a datalogger capable of measuring peak high and peak low Box Contents Pressure Vacuum Transducer Module 1 4 in male flare to 1 4 in female pipe thread connector 9 V alkaline battery Instruction Sheet Quick Reference Guide Safety Information Warning To avoid eye injury always wear approved safety glasses when working with this instrument To avoid injury from pressure exp...

Page 2: ...sured may result Cleaning Clean the module using a soft cloth and mild soap and water Do not immerse the PV350 in liquids Clean the transducer and connectors after each use with a cleaner appropriate for the liquid or gas connected during measurement Refer to the manufacturer s instructions for the proper cleaning material Measuring Pressure and Vacuum Caution The sensor diaphragm can be damaged b...

Page 3: ...le s specified accuracy 6 Apply pressure vacuum to the PV350 and read the DMM display When measuring pressure if the DMM displays OL switch the multimeter range from millivolts to dc volts Increasing the range to dc volts moves the decimal position three places to the left 500 psi displays as 0 500 V When measuring vacuum the DMM displays a minus sign Testing the Battery 1 Turn the DMM and the mod...

Page 4: ...RANGE HOLD H mA A mV V V OFF A mA A 10A MAX FUSED ZERO SMOOTH Hz RPM DUTY mA A RPM ALERT TRIGGER _ ms PULSE COM 1000V MAX 400mA MAX FUSED DC mV 4000 mV 20 2 4 6 3 8 2 4 6 8 4 V RPM METRIC ENGLISH Z PV500 PRESSURE VACUUM MODULE cmHg inHg kPa psi OFF Figure 3 Pressure for HVAC R Super Heat Measurements 88 AUTOMOTIVE TESTER MIN MAX RANGE HOLD H mA A mV V V OFF A mA A 10A MAX FUSED ZERO SMOOTH Hz RPM ...

Page 5: ...a cm Hg or in Hg Resolution Determined by the resolution of the measuring device RFI Specification Typically 1 error from 80 to 200 MHz at 3 V m Electromagnetic Compatibility EMC International IEC 61326 1 Portable IEC 61326 2 2 CISPR 11 Group 1 Class A Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequency energy that is necessary for the internal function o...

Page 6: ...Standard 9 V Alkaline battery NEDA 1604 IEC 6LR61 Battery Life 300 hours typical Electrical Cable Black PVC will withstand up to 105 C Cable Length 8 ft Pressure Port Connection 1 4 in NPT Connector Brass 1 4 in male flare to 1 4 in female pipe thread List of Replaceable Parts Part Number Description 614487 2813197 2816172 913207 926873 926881 Battery Primary 9 V 0 15 mA Case Bottom PV350 Case Top...

Reviews: