background image

P5514, P5514-V & P5514-EP 

Users Manual 

Operation 

W

 Warning 

Do not exceed the maximum working pressure of the 
instrument (stated on the instrument label). 

1.

 

Turn the screw press in to generate pressure. 

2.

 

Allow a few moments for the system to stabilize before taking any readings, 

especially after large changes in system pressure. 

Note  

Large, sudden changes in pressure will cause the system temperature to 
rise or fall, which can cause instrument readings to change as the fluid in 
the system expands or contracts, thus increasing or decreasing the 
pressure. 

3.

 

Compare the reading of the gauge under test with that of the master gauge. 

4.

 

For the next higher calibration point, repeat from Priming, Step 1. 

5.

 

To measure reducing pressures, turn the screw press out (counter-clockwise). 

6.

 

Depressurize the system by turning the screw press FULLY OUT, then open the 

reservoir valve 

W

 Caution 

Never release the system pressure by using the reservoir valve 

alone, as sudden depressurization will “shock” the system, 

which may cause damage to sensitive instruments attached 

to it. 

Maintenance and Servicing 

The Figures on the following pages detail the components of each assembly, together 

with the relevant part numbers. Where “ASSY” appears as a part number, this indicates 

that this particular component is associated with other components in an assembly for 

replacement purposes. 
Before beginning any maintenance, remove any instruments that may be mounted to the 

test pump, and drain the fluid from the system. 

Table 4. Seal Kit - Part Numbers 

Model 

Seal Kit P/N 

P5514 3891113 

P5514-V 3891124 

P5514-EP 3891136 

Screw Press Seals 

1.

 

Turn the screw press out so that there is a distance of at least 1” / 2.5cm between the 

large union nut at the end of the barrel and the capstan hub. 

2.

 

Unscrew the union nut and withdraw the lead screw assembly (24) from the barrel, 

taking care not to drop the rambler (22). 

3.

 

Inspect the seals for signs of wear or damage, replace as necessary. 

Summary of Contents for P5500 Series

Page 1: ...uke Corporation All rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies P5514 P5514 V P5514 EP Hydraulic C...

Page 2: ...vice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no...

Page 3: ...ressed Liquid 2 Personal Protective Equipment 2 Symbols Used in this Manual 2 Connections 3 Connect the Gauge 4 Connect the Test Port Insert 4 Operation 5 Priming 5 Operation 6 Maintenance and Servici...

Page 4: ...P5514 P5514 V P5514 EP Users Manual ii...

Page 5: ...iii List of Tables Table Title Page 1 Operating Fluid Compatibility 2 2 Symbols 3 3 Test Port Insert Parts List 5 4 Seal Kit Part Numbers 6 5 Screw Press Assembly Parts List 10...

Page 6: ...P5514 P5514 V P5514 EP Users Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 Connecting the Gauge 4 2 Test Port Insert 5 3 Reservoir and Test Port Assemblies 9 4 Screw Press Assembly 9 5 Angle Adapter 11 6 Pointer Remover Punch 11...

Page 8: ...P5514 P5514 V P5514 EP Users Manual vi...

Page 9: ...acy is maintained The comparison test pump should be mounted securely to a stable workbench or similar surface Four mounting holes are provided in the pump stand for this purpose How to Contact Fluke...

Page 10: ...ns that are not related to any specific procedures and do not appear elsewhere in this publication These are recommended precautions that personnel must understand and apply during equipment operation...

Page 11: ...fluids and protect the system from particulates we recommend the use of a Liquid to Liquid Separator see Section 5 Ancillary Equipment W Warning DO NOT use Teflon PTFE tape on these connections as th...

Page 12: ...the position to face forward hold the gauge adapter and turn the instrument COUNTER CLOCKWISE so that it faces forward 4 Hold the instrument steady while turning the gauge adapter COUNTER CLOCKWISE un...

Page 13: ...Open reservoir valve 13 one turn counter clockwise and turn screw press fully in 2 Close valve 13 and turn screw press fully out 3 Open valve and turn screw press fully in Note During this operation b...

Page 14: ...ution Never release the system pressure by using the reservoir valve alone as sudden depressurization will shock the system which may cause damage to sensitive instruments attached to it Maintenance a...

Page 15: ...t in the barrel 11 Re tighten the barrel union nut W Caution If the lead screw assembly 24 shows signs of excessive wear then it is very likely that the associated components have worn also therefore...

Page 16: ...n the barrel is worn also If the bore is worn or scratched it will not seal correctly and may leak under pressure 11 Re assembly is the reverse of the above Ensure that all seals and sealing surfaces...

Page 17: ...Hydraulic Comparison Test Pump Maintenance and Servicing 9 20 22 23 24 25 25 26 26 21 16 17 19 18 gjq004 eps Figure 4 Screw Press Assembly...

Page 18: ...lve Body 3923785 7 Bonded Seal 3862349 3922096 3922043 8 Reservoir 3919463 9 Locknut 3918244 10 Reservoir Cover 3918350 11 Washer 3916458 12 Spring 3922786 13 Valve Stem Assembly 3885982 14 Screw 3909...

Page 19: ...ould be used The angle adapter fits directly onto the test station converting it through 90 degrees allowing the standard adapters to be used gjn022 bmp Figure 5 Angle Adapter Pointer Remover Punch P5...

Page 20: ...P5514 P5514 V P5514 EP Users Manual 12...

Reviews: