Using the Loss/Length Option
Finding Connections with FindFiber
7-39
7
Finding Connections with FindFiber
The FindFiber function is available with the loss/length test
option. This function helps you trace connections at patch
panels and quickly check fiber continuity.
Note
The FindFiber function is not available in Far End
Source mode.
Using FindFiber in Smart Remote Mode
Use FindFiber in Smart Remote mode to help you determine
which fibers go to which connectors at a patch panel.
Figure 7-14 shows the equipment needed for using
FindFiber in Smart Remote mode.
Figure 7-14. Equipment for Using FindFiber in Smart Remote Mode
ajt50f.eps
A
Tester with OFTM-5612B or OFTM-5732 module and
smart remote (DTX remote shown). Match connector
adapters to the connectors in the link.
B
Spare battery pack (optional)
C
Two ac adapters with line cords (optional)
D
Two duplex reference test cords. Match the fiber to be
tested. One connector on each cord must be SC. For the
others match the connectors in the link.
E
Fiber cleaning supplies
Summary of Contents for OF-500-01 OptiFiber
Page 4: ......
Page 18: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook xiv...
Page 29: ...Getting Acquainted Powering the Tester 1 9 1 ajt20f eps Figure 1 1 Battery Pack Feature...
Page 31: ...Getting Acquainted Verifying Operation 1 11 1 ajt56f eps Figure 1 2 Removing the Module...
Page 48: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28...
Page 76: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 2 28...
Page 110: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 34...
Page 132: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 56...
Page 138: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6...
Page 144: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 5 6...
Page 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 44...
Page 216: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 10 10...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 12 30...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook A 2...
Page 268: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2...