Product Maintenance
The subsequent sections tell you how to maintain the Product.
Clean the Product
Caution
To prevent possible damage to the Product or to equipment
under test, remove the input signals before you clean the
Product.
Outside
To clean the outside of the Product, disconnect from the power supply and use only a
damp cloth with mild detergent.
Inside
It is possible that microbial growth can become present in the transducers of the Product.
It is recommended that you clean the transducers at 3 month intervals. To clean the inside
of the Product, inject 20 ml of a warm water and detergent solution into the Fluid Inlet
Port. After 5 minutes, flush with clean water. Always pass water from the fluid inlets to
the outlets.
Test Fluid
The Product is intended to operate with de-ionized water with added detergent. Fluids
intended for use on patients, such as high viscosity, oily, or corrosive substances will
damage the measurement system. Tap water can contain contaminates which can also
damage the transducer.
Test fluid can be made with de-ionized water and a wetting agent such as MICRO-90. It
is recommended that a 0.1 % solution of MICRO-90 in de-ionized water (preferably
degassed) be prepared in volume for daily use and kept in a sealed vessel. If the water
makes too much foam, then a 0.05 % dilution is recommended.
MICRO-90 is available from:
International Product Corp.
201 Connecticut Dr.
P.O. Box 70
Burlington, NJ 08016-0070 USA
Tel 609 386 8770
International Product Corp.
1 Church Row
Chistlehurst, Kent BR7 5PG United Kingdom
Tel. 0208 467 8944