If the Current Clamp Does Not Work
Check
Correct
Battery = OK?
Refer to “Testing the Battery.”
Voltmeter connections?
Red to + or
V
Black to
C
Voltmeter function/range?
mV dc
mV ac (or V ac)
Storage
During longer periods of non-use (> 60 days), remove and
store the battery separately.
Cleaning
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Service and Parts
The Current Clamp should be serviced only by a qualified
service technician. For service information, contact your
nearest Fluke dealer or service center.
Specifications
Operating Temperature
-10 to 50°C (14 to 122°F)
Storage Temperature
-20 to 60°C (-4 to 140°F)
Relative Humidity
0 to 95% (0 to 30
°
C)
0 to 75% (30 to 40°C)
0 to 45% (40 to 50°C)
Altitude
Operating
Non-operating
0 - 2000m
0 - 12000m
Battery Type
9V (alkaline)
Battery Life
60 hrs typical (continuous with alkaline)
Safety
s
Double Insulation, 600V rms, CAT. III
Current x Hz Product
≤
240,000
Output Signal
1 mV per amp dc or ac
Working Voltage
600V rms, CAT. III maximum at input
Maximum Conductor Size
1 ea. 30 mm (1.18 in.) diameter
2 ea. 25 mm (0.98 in.) diameter
Load Impedance
≥
1 M
Ω
,
≤
100 pF
Temperature Coefficient
+/-(0.05 x accuracy per °C (0 -18°C, 28-
50°C)
Size
209 x 78 x 48 mm
(3.09 x 8.21 x 1.87 in.)
Weight
0.5 kg (17.6 oz.)