i410/i1010

AC/DC Current Clamp

Instruction Sheet

P

 

Read First: Safety Information

  Avoid working alone to ensure that assistance can be rendered.

  Do not use the clamp if the clamp or leads appear damaged, or if you

suspect that the clamp is not operating properly.

  Use extreme caution when working around bare conductors or bus

bars. Accidental contact with the conductor could result in electric
shock.

  Use caution when working with voltages above 30V rms or 60V dc.

Such voltages pose a shock hazard.

  Never use the clamp on a circuit with a voltage higher than 600V

(CAT. III) or a frequency higher than 400 Hz (i410) or 2 kHz (i1010.)
The clamp may be damaged if these limits are exceeded.

  Use the clamp only as specified in this instruction sheet; otherwise,

the protection provided by the clamp may be impaired.

  Read the operating instructions before use and follow all safety

information.

  CAT. III - Industrial Locations (IEC 1010-1).

PN 617677  September 1996, Rev.1, 4/97

1996, 1997 Fluke Corporation. All rights reserved. Printed in Thailand.

Summary of Contents for i410

Page 1: ...n when working with voltages above 30V rms or 60V dc Such voltages pose a shock hazard Never use the clamp on a circuit with a voltage higher than 600V CAT III or a frequency higher than 400 Hz i410 or 2 kHz i1010 The clamp may be damaged if these limits are exceeded Use the clamp only as specified in this instruction sheet otherwise the protection provided by the clamp may be impaired Read the op...

Page 2: ...RMS MULTIMETER ar001i eps A Jaw Lever B Jaw Centering Marks C ON Indicator D ON OFF Switch E Zero Adjust F Battery Access G Output Cable Minimum voltmeter requirements Accepts safety shrouded banana plugs Can display 1 mV 0 1 mV preferred Accuracy 0 75 Input impedance 1 MΩ 100 pF ...

Page 3: ...METER i410 ZERO A 600V CAT 600V 600A CLAMP METER i410 ZERO A CAT 600V 600A 600V mV 76 TRUE RMS MULTIMETER ar002i eps A Connect to voltmeter B Select mV dc C Set ON D Adjust ZERO jaws empty ZERO A ar006i eps E Clamp and center around conductor F Read voltmeter 1 mV 1A ...

Page 4: ...0 Speed CLAMP METER i410 ZERO A 600 V CAT 600V 600A CLA MP ME TER i4 10 ZE RO A 60 0V CA T 60 0V 60 0A V 76 TRUE RMS MULTIMETER ar003i eps A Connect to voltmeter B Select mV ac or V ac but resolution may be limited to 1A C Set ON D Clamp and center around conductor E Read voltmeter 1 mV 1A ...

Page 5: ...10 ZERO A CAT 600V 600A 600V V 76 TRUE RMS MULTIMETER ar004i eps A Set OFF B Select V dc C 7 0V dc replace battery Voltmeter input impedance 1 MΩ Replacing the Battery For specified battery life use an alkaline battery 1 2 3 ar005i eps ...

Page 6: ...dealer or service center Specifications Operating Temperature 10 to 50 C 14 to 122 F Storage Temperature 20 to 60 C 4 to 140 F Relative Humidity 0 to 95 0 to 30 C 0 to 75 30 to 40 C 0 to 45 40 to 50 C Altitude Operating Non operating 0 2000m 0 12000m Battery Type 9V alkaline Battery Life 60 hrs typical continuous with alkaline Safety s Double Insulation 600V rms CAT III Current x Hz Product 240 00...

Page 7: ... Accuracy zero adjusted conductor centered 3 5 0 5A 0 400A 2 0 0 5A 0 1000A AC Accuracy 3 5 0 5A 45 400 Hz Crest Factor 3 0 400A 2 0 0 5A 45 400 Hz Crest Factor 3 3 0 0 5A 400 Hz 2 kHz sine wave 0 600A Bandwidth 3 kHz 10 kHz With a true rms voltmeter the minimum ac current is limited to the low end of the specified mV ac range ...

Page 8: ...the warranty period To obtain warranty service contact your nearest Fluke authorized service center or send the product with a description of the difficulty postage and insurance prepaid FOB Destination to the nearest Fluke authorized service center Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepaid FOB Destination If ...

Reviews: