Fluke i3000s Flex Series Instruction Sheet Download Page 2

PHASE SHIFT 

30A Range 

 

Figure 4. Phase Shift @ 10A 

300A Range 

 

Figure 5. Phase Shift @ 10A 

3000A Range 

 

Figure 6. Phase Shift @ 100A 

 

Figure 7. Maximum Current as a Function of the Frequency 

 

Instrument Compatibility 

The i3000s is compatible with any Fluke ScopeMeter test tool, 
Power Harmonics Analyzer, Oscilloscope, Multimeter, or other 
voltage measurement device that has the following features: 

 

BNC input connector. The Dual Banana to BNC Adapter 
included in the package, can be used to connect to standard 
inputs on multimeters. For the 120 series ScopeMeters, use 
the BB120 Shielded Banana to BNC Adapter. 

 

Input accuracy of 2% or better to take full advantage of the 
accuracy of the Current Clamp. 

 

Input impedance of greater than or equal to 1 M

, and for full 

bandwidth and accuracy, a maximum input capacity of 47 pF. 

 

A passband of more than four times the frequency of the 
waveform to be measured. 

 

Using the Current Clamp 

To use the Current Clamp, follow these instructions: 
1. 

Connect the i3000s Current Clamp to the desired input on the 
measuring instrument. When you are using a multimeter, use 
the Dual Banana to BNC Adapter (PM9081/001) to connect the 
Current Clamp to the input. (See Figure 8.) 

2. 

On the Current Clamp, select the least sensitive range (0.1 
mV/A). 

3. 

Select the appropriate clamp sensitivity on your ScopeMeter 
test tool or oscilloscope. 

4. 

Position the Current Clamp perpendicularly and centered 
around the conductor.  
Make sure that the arrow marked on the clamp jaw points 
toward the correct orientation for correct phase display on the 
oscilloscope. (See Figure 9.)   
Use the markings on the jaws to center the conductor. 

5. 

Observe the current value and waveform on the instrument’s 
display. 

6. 

If desired, select a lower range on the Current Clamp and set 
the corresponding sensitivity (mV/A setting) on the 
ScopeMeter test tool or oscilloscope. 

 
If the corresponding sensitivity is not available on the ScopeMeter 
or oscilloscope, select the closest setting and calculate the actual 
current value from the displayed. 

Example1:

 Current Clamp set to 0.1 mV/A, ScopeMeter test tool 

set to 1 mV/A. 
The ScopeMeter test tool displays 50A.  
The real current value can be calculated with the following formula: 

Actual current

display value

sensitivity instrument

sensitivity Current Clamp

mV

A

mV

A

A

=

×

=

×

=

50

1

01

500

/

.

/

 

 

Example 2:

 Current Clamp set to 10 mV/A. Multimeter displays 

0.285V. 

Actual current

display value

sensitivity Current Clamp

V

mV A

mV

mV A

A

=

=

=

=

0 285

10

285

10

28 5

.

/

/

.

 

V

COM

V

ST8386_190

 

Figure 8. Measurement Setup 

 

 Warning 

If the sensitivity setting (mV/A) of the ScopeMeter test 
tool or oscilloscope does not correspond with the 
setting of the Current Clamp, the instrument may 
display a much lower current than the actual value. 
This may cause misinterpretation and as a 
consequence incorrect handling.  

 

 

Figure 9. Orientation of the Current Clamp 

 

Measurement Considerations 

Observe the following guidelines for positioning the Current Clamp 
Jaws: 

 

Center the conductor inside the clamp jaws. 

 

Make sure the clamp is perpendicular to the conductor. 

 

Make sure that the arrow marked on the jaw of the Current 
Clamp points toward the correct direction. 

Observe the following guidelines when making measurements: 

 

If possible, avoid measurements close to other current-carrying 
conductors. 

 

On the Current Clamp, select the most appropriate range for 
the current being measured to get the best accuracy. 

 

Maintenance 

Before each use, assure continued safety by inspecting the clamp. 
Look for cracks or missing portions of the clamp housing and output 
cable insulating cover and for loose or weakened components. Pay 
particular attention to the insulation surrounding the clamp jaws. If a 
clamp fails this inspection, tape it shut to prevent unintended 
operation. A clamp under warranty will be promptly repaired or 
replaced (at Fluke's discretion) and returned at no charge. 

Cleaning and Storage 

Periodically wipe the case with a damp cloth and detergent; do not 
use abrasives or solvents. Open the jaws and wipe the magnetic 
pole pieces with a lightly oiled cloth. Do not allow rust or corrosion 
to form on the magnetic core ends. 
 
 

If your Current Clamp does not work 

If the Current Clamp does not perform properly, use the following 
steps to help isolate the problem: 

 

Inspect the jaw mating surface for cleanliness. If any foreign 
material is present, the jaws will not close properly and errors 
will result. 

 

Verify that the function selection on the ScopeMeter test tool or 
oscilloscope is correct, i.e., the display vertical resolution is not 
too low or too high.  

 

LIMITED WARRANTY & LIMITATION OF LIABILITY 

This Fluke product will be free from defects in material and 
workmanship for one year from the date of purchase. This warranty 
does not cover fuses, disposable batteries or damage from 
accident, neglect, misuse or abnormal conditions of operation or 
handling. Resellers are not authorized to extend any other warranty 
on Fluke’s behalf. To obtain service during the warranty period, 
send your defective product to the nearest Fluke Authorized Service 
Center with a description of the problem. 

 

THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER 
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR 
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT 
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR 
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY 
CAUSE OR THEORY.  
Since some states or countries do not allow the exclusion or 
limitation of an implied warranty or of incidental or consequential 
damages, this limitation of liability may not apply to you. 

 

Fluke Corporation  

Fluke Industrial B.V. 

 

P.O. Box 9090 

P.O. Box 90 

 

Everett WA 

7600 AB Almelo 

 

98206-9090, USA 

The Netherlands 

 

 
 

SERVICE CENTERS

 

To locate an authorized service center,  

visit us on the World Wide Web:  

http://www.fluke.com

  

or call Fluke using any of the phone numbers listed below: 

 

+1-888-993-5853 in U.S.A. and Canada 

 

+31-40-267-5200 in Europe 

 

+1-425-446-5500 from other countries 

 

Tactile barrier 

Reviews: