Booster
Installation and Setup
7
W
Warning
To prevent personal injury, make sure that maximum input pressure to the booster
drive adjust regulator (
) is below 1.7 MPa (250 psi). Pressures above this level may
result in a failure that could damage the instrument and/or cause personal injury.
5. Connect the test gas supply to the 1/4 in. NPT female SUPPLY port (
E
) with tubing of the appropriate pressure
rating (not supplied).
W
Warning
To prevent personal injury, make sure that maximum input pressure to the SUPPLY port
(
E
) is 41 MPa (6 000 psi) as shown by the supply pressure gauge (
D
). Pressure above
this level may result in a failure that could damage the instrument and/or cause
personal injury.
Note
Pressure that is present at the SUPPLY IN port (
E
) is present at the inlet port of the high-
pressure adjust regulator (
F
). To prevent gas pressure from reaching the HIGH PRESSURE
OUT port (
H
), make sure that the high-pressure adjust regulator is closed (backed off).
6. Connect the HIGH PRESSURE OUT port (
H
) to the high pressure SUPPLY port of the Pressure Controller.
Working pressure rating of the lines and fittings should be at least 61 MPa (8850 psi) for GBK-50M and 124 MPa
(18 000 psi) for GBK-110M. The GBK-50M uses a 1/4 NPT port. The GBK-110M uses a DH500 connection
(cone and threaded connections compatible with Autoclave F250C and HIP HF4).
W
Warning
To prevent personal injury, know that using a connecting tubing with a working
pressure below the maximum pressure generated by the booster may result in a failure
in the tubing that could cause damage to the instrument and/or cause personal injury.
Summary of Contents for GBK-110M
Page 18: ......