Singlemode test reference cord: 

0.20 dB or less for 1310 nm and 1550 nm

5

If a result is more than the limit shown above, use the procedure 
given in the Users Manual to test the cords in the path that 
failed (Input Fiber or Output Fiber). When you have good test 
reference cords, set the reference and do these steps again 
before you test a fiber link.

6

Save the test results to show that you used good test reference 
cords to test the link.

Note

Fluke Networks recommends that you use this procedure 
to make sure your test reference cords are good before 
you test each fiber link.

gfa03.eps

Figure 3. Testing the Test Reference Cords in 

Smart Remote Mode

Summary of Contents for DTX-CLT CertiFiber

Page 1: ...t versions of the DTX CLT CertiFiber Users Manual and the DTX Series Technical Reference Handbook provided on the DTX CLT CertiFiber Product Manuals CD and on the Fluke Networks website WWarningX To avoid possible fire electric shock personal injury damage to equipment or inaccurate test results read the safety information in the DTX CLT CertiFiber Users Manual before using the tester A Power the ...

Page 2: ...select Current Folder Select an existing folder or press J Create Folder to create a new folder Enter job information On the Instrument Settings menu press C to show the tab with the Operator Site and Company names To enter a new name select a setting press J Create then use the softkeys BC A D and H for editing Press N when you are finished C Clean and Inspect All Connectors WCaution When using t...

Page 3: ...ress Cto see other tabs Test Limit Select the test limit required for the job Press J More to see other lists of limits Fiber Type Select the fiber type to be tested Remote End Setup Set to Smart Remote Bi directional Enable this if you are required to test the fiber in both directions Number of Adapters and Number of Splices Enter the number of adapters and splices that will be added to each dire...

Page 4: ...tch to AUTOTEST then press P 4 Press H to see the test results The results for each fiber Input Fiber and Output Fiber must be no more than these limits WCaution An overall result of PASS does not show that the test reference cords are good You must compare the loss to the limits given here to make sure the cords are good Multimode test reference cord 850 nm 0 11 dB or less 1300 nm 0 10 dB or less...

Page 5: ...utput Fiber When you have good test reference cords set the reference and do these steps again before you test a fiber link 6 Save the test results to show that you used good test reference cords to test the link Note Fluke Networks recommends that you use this procedure to make sure your test reference cords are good before you test each fiber link gfa03 eps Figure 3 Testing the Test Reference Co...

Page 6: ...o a Test 1 Connect to the fiber you will test Figure 4 shows Smart Remote mode 2 Turn the rotary switch to AUTOTEST then press P gda04 eps Figure 4 Testing a Fiber Link in Smart Remote Mode 1 Jumper Method ...

Page 7: ...te at www flukenetworks com registration Contacting Fluke Networks www flukenetworks com support flukenetworks com 1 425 446 4519 Australia 61 2 8850 3333 or 61 3 9329 0244 Beijing 86 10 6512 3435 Brazil 11 3759 7600 Canada 1 800 363 5853 Europe 31 0 40 2675 600 Hong Kong 852 2721 3228 Japan 03 6714 3117 Korea 82 2 539 6311 Singapore 65 6799 5566 Taiwan 886 2 227 83199 USA 1 800 283 5853 Visit our...

Page 8: ...nty on Fluke Networks behalf To obtain service during the warranty period contact your nearest Fluke Networks authorized service center to obtain return authorization information then send your defective product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE NETWO...

Reviews: