DTX Series CableAnalyzer
Technical Reference Handbook
1-24
Verifying Operation
The tester performs a basic self test when you turn it on. To
run a more thorough self test for an acceptance test or as
part of a routine equipment check:
1
Connect the main and remote testers using one of the
methods shown in Figure 1-10.
2
Turn the rotary switch to SPECIAL FUNCTIONS.
3
Use
D
to highlight Self Test; then press
H
.
4
If fiber modules are installed in addition to twisted pair
adapters, select Mainframe.
5
Press
P
.
If the tester reports an error, refer to “If Something Seems
Wrong” on page 14-9.
amd168.eps
Figure 1-10. Self Test Connections
Permanent link
adapter
Channel
adapter
TEST
LOW BATTERY
TONE
TALK
FAIL
TEST
PASS
TALK
TALK
MONITOR
ENTER
TEST
SAVE
SPECIAL
FUNCTIONS
SETUP
AUTO
TEST
SINGLE
TEST
EXIT
F1
F2
F3
TE
S
T
LOW BA
TTE
R
Y
TONE
TA
L
K
FA
IL
TE
S
T
PA
SS
TA
LK
TA
LK
MONITOR
ENTER
TE
S
T
S
AV
E
S
PECIAL
FUNCTION
S
S
ETUP
AUTO
TE
S
T
S
IN
G
LE
TE
S
T
EXIT
F1
F2
F3
DTX-REFERENCE module
Fiber modules
Twisted pair adapters
Summary of Contents for DTX-1800
Page 4: ......
Page 18: ...DTX Series CableAnalyzer Technical Reference Handbook xiv ...
Page 132: ...DTX Series CableAnalyzer Technical Reference Handbook 4 8 ...
Page 158: ...DTX Series CableAnalyzer Technical Reference Handbook 6 14 ...
Page 270: ...DTX Series CableAnalyzer Technical Reference Handbook 8 62 ...
Page 284: ...DTX Series CableAnalyzer Technical Reference Handbook 10 10 ...
Page 304: ...DTX Series CableAnalyzer Technical Reference Handbook 11 20 ...
Page 372: ...DTX Series CableAnalyzer Technical Reference Handbook A 8 ...
Page 374: ...DTX Series CableAnalyzer Technical Reference Handbook B 2 ...