Certifying Fiber Optic Cabling
Using the Power Meter
8-57
8
Using the Power Meter
The power meter lets you measure the optical power
produced by a source such as an optical network interface
card or optical test equipment.
The tester offers two versions of the power meter function:
•
SINGLE TEST mode: Takes one power measurement at
850 nm and 1300 nm (DTX-MFM2), 850 nm and
1310 nm (DTX-GFM2), or 1310 nm and 1550 nm
(DTX-SFM2). You can save the power measurement in
this mode.
•
MONITOR mode: Monitors power continuously at the
input port at 850 nm, 1300 nm, 1310 nm, or
1550 nm. This measurement cannot be saved.
The power meter functions do not compare the power
measurement to a limit and do not produce PASS/FAIL
results.
Figure 8-29 shows the equipment required for using the
power meter in MONITOR mode.
Figure 8-29. Equipment for Using the Power Meter in
MONITOR Mode
amd57f.eps
A
Tester with fiber module and connector adapter
installed. Match the connector adapter to the
connector in the link or on the source. See
B
Memory card (optional for SINGLE TEST mode)
C
AC adapter with line cord (optional)
D
One test reference cord. Match the fiber used in the
installation, with SC at tester end. See Figure 8-30.
E
Fiber cleaning supplies
C
A
D
2
5
Summary of Contents for DTX-1800
Page 4: ......
Page 18: ...DTX Series CableAnalyzer Technical Reference Handbook xiv ...
Page 132: ...DTX Series CableAnalyzer Technical Reference Handbook 4 8 ...
Page 158: ...DTX Series CableAnalyzer Technical Reference Handbook 6 14 ...
Page 270: ...DTX Series CableAnalyzer Technical Reference Handbook 8 62 ...
Page 284: ...DTX Series CableAnalyzer Technical Reference Handbook 10 10 ...
Page 304: ...DTX Series CableAnalyzer Technical Reference Handbook 11 20 ...
Page 372: ...DTX Series CableAnalyzer Technical Reference Handbook A 8 ...
Page 374: ...DTX Series CableAnalyzer Technical Reference Handbook B 2 ...