Wireless K-Type Thermometer
Required Equipment
5
Required Equipment
The equipment in Table 2 is necessary for performance tests and calibration adjustment.
Table 2. Required Equipment
Equipment Required
Characteristics
Recommended
Model
Notes
Calibrator
Output Voltage: 30 mV dc
Accuracy: ±0.125 %
Resolution: 0.1
µ
A
Fluke 5500A Calibrator
(or equivalent)
Calibration only
Temperature Accuracy
Verification only
-25 to 100
°
C
-200 to 1300
°
C
±0.16
°
C
±0.40
°
C
Copper Wire
Calibration only
Thermocouple
Connector
Copper
Fluke PN 601747
Calibration only
Thermocouple
Connector
For K-type thermocouple
Fluke 80CK-M
PN 779942
Verification only
K-type
Thermocouple
Wire
Standard Omega
Verification
only
Use copper wire, not thermocouple wire, to source voltage during calibration. Use copper plugs with
copper wire. Use yellow K-type plugs and wires to source temperature.