Fluke Cable Analyzer DTX-ELT Brochure & Specs Download Page 3

Power

Main unit and remote: Lithium-ion battery pack, 7.4 V, 4000 mAh
  • Typical battery life: 12 to 14 hours
  • Charge time (with tester off): 4 hours (below 40 °C)
  • AC adapter/charger, USA version: Linear power supply; Input 108 V ac to 132 V ac, 60 Hz ; 
     Output 15 V dc, 1.2 A
  • AC adapter/charger, international version: Switching power supply; Input 90 V ac to 264 V ac, 
     48 to 62 Hz; Output 15 V dc, 1.2 A (isolated output)
  • Memory backup power in main unit: Lithium battery
  • Typical life of lithium battery: 5 years
  • The battery will not charge at temperatures outside of 0°C to 45°C (32°F to 113°F). 
     The battery charges at a reduced rate between 40°C and 45°C (104°F and 113°F). 

Languages Supported

English, French, German, Spanish, Russian and Italian

Calibration

Service center calibration period is 1 year

Performance Specifications

Note:

 all specifications for tests on twisted-pair cabling apply to 100 

Ω

   cable. Contact Fluke Networks 

for information on measurement performance for cable with a different impedance.

Baseline accuracy

The baseline accuracy of all DTX Series testers meets or exceeds the requirements of Level IV in 

 

2nd Edition of IEC 61935-1 over the frequency range supported by the tester.

Category 6/Class E link  

test modes (and lower link  

categories)

DTX CableAnalyzer far exceeds Level III requirements of TIA/EIA-568-B.2-2 and IEC 61935-1

Length for Twisted Pair Cabling

(Length specs do not include the uncertainty of the cable’s NVP Value)

Parameter

Single-ended Test

Dual-ended Test

 (Main and Remote)

Range

800 m (2600 ft) 

150 m (490 ft)

Resolution

0.1 m or 1 ft 

0.1 m or 1 ft 

Accuracy

± (1m + 4%) 

± (1m + 4%)

DC Loop Resistance Test
Parameter

Twisted Pair Cabling

Range

Ω

  to 53

Ω

Resolution

0.1 

Ω

Accuracy

± (1 

Ω

 + 1%)

Overload Recovery Time

Less than 10 minutes to rated accuracy following over-voltage. Referencing is required after repeated or 
prolonged over-voltage.

Delay Skew
Parameter

Twisted Pair Cabling

Range

0 ns to 100 ns

Resolution

1 ns

Accuracy

± 10 ns