![Fluke BT510 User Manual Download Page 22](http://html1.mh-extra.com/html/fluke/bt510/bt510_user-manual_2309828022.webp)
BT520
Users Manual
1-12
Accuracy Specifications
Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F
to 82 °F), with relative humidity to 80 %. Accuracy specifications are given as: ±([% of
reading] + [number of least significant digits]). Accuracy specification assumes ambient
temperature stable ±1 °C.
Function Range
Resolution
Accuracy
Battery Internal Resistance
[1]
3 m
Ω
0.001
m
Ω
1 % + 8
30 m
Ω
0.01
m
Ω
0.8 % + 6
300 m
Ω
0.1
m
Ω
0.8 % + 6
3000 m
Ω
1
m
Ω
0.8 % + 6
V dc
6 V
0.001 V
0.09 % + 5
60 V
0.01 V
600 V
0.1 V
V ac (45 Hz to 500 Hz with low-pass
filter)
600 V
0.1 V
2 % + 10
Frequency (Display with V ac)
Trigger level:
≥
10 mV @V ac;
45 Hz to 500 Hz
0.1 Hz
0.5 % + 8
AC Voltage Ripple (20 kHz max)
600 mV
6000 mV
0.1 mV
1 mV
3 % + 20
3 % + 10
[1] The measurement is based on ac injection method. The injected source signal is <100 mA, 1 kHz.
Summary of Contents for BT510
Page 6: ...BT520 Users Manual iv...
Page 8: ...BT520 Users Manual vi...
Page 10: ...BT520 Users Manual viii...
Page 24: ...BT520 Users Manual 1 14...
Page 48: ...BT520 Users Manual 3 18...
Page 66: ...BT520 Users Manual 7 6 hsn032 eps Figure 7 3 Charge the Battery...
Page 68: ...BT520 Users Manual 7 8...