5.4
Right Side Panel
The right side panel is illustrated in Figure 2.
Test Well
- The right side panel has a 6.3 mm (0.25") hole near its center that
allows access to a 3.18 mm (0.125") test well in the IR target plate. The test
well accepts 3.18 mm (0.125") RTDs, thermistor probes, or thermocouples for
calibration (see Section 8.2). The test well is also used to calibrate the 9135
against a standard RTD (see Section 9).
Calibration Access Hole
- There is a small oblong hole on the right side panel.
These gives access to two potentiometers used to calibrate the 9135. DO NOT