7 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 9999.99 {C or F}
set: 150.00 C
Set current set-point to n
s[etpoint]=n
s=450
Instrument
Range
Read scan function
sc[an]
sc
scan: {ON or OFF}
scan: ON
Set scan function
:
sc[an]=on/of[f]
ON or OFF
Turn scan function on
sc[an]=on
sc=on
Turn scan function off
sc[an]=of[f]
sc-of
Read scan rate
sr[ate]
sr
srat: 999.99 {C or F}/min srat: 10.0 C/min
Set scan rate to
n
degrees per
minute
sr[ate]=
n
sr=5
.1 to 100°C
Read temperature
t[emperature]
t
t: 9999.99 {C or F}
t: 55.69 C
Secondary Menu
Read proportional band setting
pr[op-band]
pr
pb: 999.9
pr: 15.9
Set proportional band to
n
pr[op-band]=n
pr=8.83
Depends on
Configuration
Read cutout setting
c[utout]
c
c: 9999 {C or F}
c: 620 C, in
Set cutout setting:
c[utout]=n/r[eset]
Set cutout to
n
degrees
c[utout]=n
c=500
Temperature
Range
Reset cutout now
c[utout]=r[eset]
c=r
Read heater power
(duty cycle)
po[wer]
po
p%: 9999
po: 1
Ramp and Soak Menu
Read number of programmable
set-points
pn
pn
pn: 9
pn: 2
Set number of programmable
set-points to
n
pn=
n
pn=4
1 to 8
Read programmable set-point
number
n
ps
n
ps3
ps
n
: 9999.99 {C or F}
ps1:
50.00 C
Set programmable set-point num-
ber
n
to
n
ps
n
=
n
ps3=50
1 to 8, Instru-
ment Range
Read program set-point soak
time
pt
pt
ti: 999
ti: 5
Set program set-point soak time
to
n
minutes
pt=
n
pt=5
0 to 500
Read program control mode
pc
pc
prog: {OFF or ON}
prog: OFF
Table 3.
Digital Interface Command Summary