Preparing the Meter for Operation
Power Considerations
2
2-5
Table 2-1. Line Voltage to Fuse Rating
Line Voltage Selection
Fuse Rating
Fluke Part No.
100
0.25 A, 250 V (slow blow)
166306
120
0.25 A, 250 V (slow blow)
166306
220
0.125 A, 250 V (slow blow)
166488
240
0.125 A, 250 V (slow blow)
166488
120
caw0201f.eps
Figure 2-1. Replacing the Line Fuse
Current-Input Fuses
The 400 mA and 10 A inputs are protected by user-replaceable fuses.
•
The 400 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 943121).
•
The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 803293).
W
Warning
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
To test for a blown Current Input Fuse:
1.
With the Meter powered up, plug a test lead into the
V
Ω
GER
connector.
2.
Press
N
.
3.
Insert the other end of the test lead into the 400 mA input connector.
If the fuse is good, the Meter will read less than 200
Ω
. If the fuse is blown, the
Meter will read
overload
.
4.
Remove the probe from the 400 mA connector and insert into the 10 A
connector.
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...