Calibration
25
Table 5. Calibration Inputs (cont.)
Step
Value
Frequency
Time to Settle (In Seconds)
AC V
1
5 V
700 Hz
7
2
50 V
700 Hz
7
3
500 V
700 Hz
7
4
500 V
700 Hz
7
DC V
1
5V
DC
7
2
50 V
DC
7
3
500 V
DC
7
4
500 V
DC
7
Ohms
1
1
2
0 Ohms
16
2
500 Ohms
16
3
5k Ohms
7
Ohms
4
50k Ohms
7
5
500k Ohms
7
6
5M Ohms
7
3
7
30M ohms
1
3
AC uA
1
500 uA
700 Hz
7
2
5000 uA
700 Hz
7
DC uA
1
500 uA
DC
16
2
5000 uA
DC
7
AC mA
1
50 mA
700 Hz
7
2
400 mA
700 Hz
7
DC mA
1
50 mA
DC
16
2
400 mA
DC
7
1. Use 5520A 2-Wire Comp.
2. Zero ohms must use the same calibrator and test leads as all other ohm cal points.
Do not bypass the calibrator with a shorting bar for the zero ohms step.
3. Because of capacitive coupling, readings will be affected by movement of the leads or
operator. Put fingers on UP button and wait for display reading to settle before pushing.