calibrator signal generator for 1V rms at 25 MHz and the 
DMM to measure 1V dc. 

C1

C1

220 pf

R1

220 k

R2

500 k

AC INPUT

DC OUTPUT

CR1

CR1

CR2

CR2

R1

R3

R3
1.3 M

R2

 

Figure 1. Schematic Diagram 

1

2

5

4

3

Fluke 85RF Probe under test

6

 

Required Equipment List 

The following equipment (or equivalent) is required for 
testing the performance of the 85RF:  

EQUIPMENT TYPE 

RECOMMENDED MODEL 

A

 

AC Calibrator 

Fluke 5520A-SC600 

B

 

50-Ohm Feed Thru 
Termination  

 
Fluke Y9103 

C

 

BNC Jack to Banana 
Plug Adapter 

 
Fluke Y9108 

D

 

BNC to Probe Adapter 

Fluke P/N 574756 

E

 

Digital Multimeter, 3-
1/2 Digit (DMM) 

 
Fluke 87-III 

F

 

BNC Jack to Jack 
Adapter 

 
Pomona 3283 

Figure 2. 

Equipment Configuration for Performance Testing 

2.

 

The DMM should read between 0.950  and 1.050V dc. If 

the probe is functioning but is not within these limits, access 
and adjust R2 for a DMM reading between 0.990 and 1.010V 
dc. Reassemble the probe. 
3.

 

Transfer the unit under test to the Normal output of the 

calibrator and reduce the output level to 0.250V rms 1 MHz. 
The DMM should read between 0.223 and 0.280V dc. 
4.

 

Set the AC Calibrator to STDBY, decrease the 

frequency to 100 kHz, and increase the voltage to 30V ac 
rms. (Do not exceed this value.) 