789
Users Manual
26
Table 9. mA Stepping Pushbuttons
Pushbutton Adjustment
X
M
% STEP
Adjusts up to the next higher 25 %
step
% STEP
G
W
Adjusts down to the next lower 25
% step
F
Span Check
Sets to 100 % value
Span Check
E
Sets to 0 % value
Table 10. mA Step Values
Value (for each span setting)
Step
4 to 20 mA
0 to 20 mA
0 %
4.000 mA
0.000 mA
25 %
8.000 mA
5.000 mA
50 %
12.000 mA
10.000 mA
75 %
16.000 mA
15.000 mA
100 %
20.000 mA
20.000 mA
125 %
24.000 mA
120 %
24.000
mA
Auto Ramping the mA Output
Auto ramping gives the ability to continuously apply a
varying current stimulus from the meter to a transmitter,
while hands remain free to test the response of the
transmitter. Select either sourcing or simulating by
choosing the SOURCE or SIMULATE jacks.