Pressure Calibrator
Specifications
5
1-year Specifications
Low Pressure Sensor
High Pressure Sensor
Model
Calibrator
Description
Range
Sensor 1
Resolution
Sensor 1
Accuracy
Sensor 1
Range Sensor 2
Resolution
Sensor 2
Accuracy
Sensor 2
721-1601
16 PSIG, 100 PSIG
-14 psi + 16 psi
-0.97 bar to
1.1 bar
0.001 psi,
0.0001 bar
0.025 % of
full scale
-12 psi to +100 psi
-0.83 bar to 6.9 bar
0.001 psi
0.0001 bar
0.025 % of
full scale
721-1603
16 PSIG, 300 PSIG
-12 psi to +300 psi
-0.83 bar to 20 bar
0.01 psi
0.001 bar
721-1605
16 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
721-1610
16 PSIG, 1000
PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.01 psi
0.001 bar
721-1615
16 PSIG, 1500
PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.1 psi
0.01 bar
721-1630
16 PSIG, 3000
PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.1 psi
0.01 bar
721-1650
16 PSIG, 5000
PSIG
0 psi to +5000 psi
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.035 % of
full scale
721-3601
36 PSIG, 100 PSIG
-14 psi + 36 psi
-0.97 bar to
2.48 bar
-12 psi to +100 psi
-0.83 bar to 6.9 bar
0.001 psi
0.0001 bar
0.025 % of
full scale
721-3603
36 PSIG, 300 PSIG
-12 psi to +300 psi
-0.83 bar to 20 bar
0.01 psi
0.001 bar
721-3605
36 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
721-3610
36 PSIG, 1000
PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.01 psi
0.001 bar
721-3615
36 PSIG, 1500
PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.1 psi
0.01 bar
721-3630
36 PSIG, 3000
PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.1 psi
0.01 bar
721-3650
36 PSIG, 5000
PSIG
0 psi to +5000 psi
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.035 % of
full scale
Electromagnetic Environment ....................... IEC 61326-1: Portable
Electromagnetic Compatibility....................... Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be used
in homes.
Summary of Contents for 721-1601
Page 4: ...721 Calibration Manual ii...
Page 6: ...721 Calibration Manual iv...
Page 7: ...v List of Figures Figure Title Page 1 Battery Replacement 17 2 User Replaceable Parts 20...
Page 8: ...721 Calibration Manual vi...
Page 25: ...Pressure Calibrator Maintenance 17 hmq061 eps Figure 1 Battery Replacement...