background image

2  Fluke Corporation      Fluke 719 Electrical Pressure Calibrator

 

Ordering information

Fluke-719

 

30G

 Electric Pressure Calibrator, 

30 psi, 2 bar

Fluke-719 

100G

 Electric Pressure  

Calibrator, 100 psi, 7 bar

Includes:

 TL75 Test leads, AC72 Alligator 

clip set, Push fit connectors, Translucent test 

hose, Product overview manual (print), and 

User’s manual (CD-ROM) in 14 languages

Pump specifications

  Maximum pressure generation of 120 PSI  

(8 BAR) (100 PSI model)

  Pump to:

-  30 PSI in less than 15 seconds*

  -  100 psi in less than 45 seconds* 
 

*(with 1 meter of 1/8 inch test hose)

  5000 pump cycles between rebuilds 

  Total pump life 20,000 calibration cycle

General specifications

Storage temperature:

 -30 °C to 60 °C

Operating temperature:

 -10 °C to 55 °C

Relative humidity:

 95 % (-10 °C to 30 °C);  

75 % (30 °C to 40 °C); 45 %  

(40 °C to 50 °C); 35 % (50 °C to 55 °C)

Shock:

 1 meter drop test

Safety:

 CSA C22.2 No. 1010.1:1992

EMC:

 EN50082-1:1992 and EN55022:1994 Class B

Size (HxWxL):

 60 mm x 87 mm x 210 mm  

(2.19 in x 3.41 in x 8.28 in)

Weight:

 912 g (2 lb) 

Battery:

 Two 9 V alkaline batteries

Battery life:

 12 hours with 12 mA into 500 

, 200 

pump cycles to 100 psi, 500 pump cycles to 30 psi

Warranty:

 Three-years, one year for pump

Specifications

Function 
Measure or Source

Range

Resolution

Accuracy

Notes

Fluke-719 30G

-12  to 36 psi,  

-850 mbar to 2.4 bar

0.001 psi, 0.1 mbar 0.025 % range for 6 

months. 
0.035 % of range for 

one (1) year

Fluke-719 100G

-12 to 120 psi,  

-850 mbar to 8 bar 

0.01 psi, 1 mbar

mA

0 to 24 mA

0.001 mA

.015 % Rdg + 2 counts Max load, 1000 

Ω 

7

50 

Ω 

ohm in HART mode 

Loop supply

24 V dc

N/A

±

 10 %

Reviews: