2 Fluke Corporation Fluke 719 Electrical Pressure Calibrator
Ordering information
Fluke-719
30G
Electric Pressure Calibrator,
30 psi, 2 bar
Fluke-719
100G
Electric Pressure
Calibrator, 100 psi, 7 bar
Includes:
TL75 Test leads, AC72 Alligator
clip set, Push fit connectors, Translucent test
hose, Product overview manual (print), and
User’s manual (CD-ROM) in 14 languages
Pump specifications
•
Maximum pressure generation of 120 PSI
(8 BAR) (100 PSI model)
•
Pump to:
- 30 PSI in less than 15 seconds*
- 100 psi in less than 45 seconds*
*(with 1 meter of 1/8 inch test hose)
•
5000 pump cycles between rebuilds
•
Total pump life 20,000 calibration cycle
General specifications
Storage temperature:
-30 °C to 60 °C
Operating temperature:
-10 °C to 55 °C
Relative humidity:
95 % (-10 °C to 30 °C);
75 % (30 °C to 40 °C); 45 %
(40 °C to 50 °C); 35 % (50 °C to 55 °C)
Shock:
1 meter drop test
Safety:
CSA C22.2 No. 1010.1:1992
EMC:
EN50082-1:1992 and EN55022:1994 Class B
Size (HxWxL):
60 mm x 87 mm x 210 mm
(2.19 in x 3.41 in x 8.28 in)
Weight:
912 g (2 lb)
Battery:
Two 9 V alkaline batteries
Battery life:
12 hours with 12 mA into 500
Ω
, 200
pump cycles to 100 psi, 500 pump cycles to 30 psi
Warranty:
Three-years, one year for pump
Specifications
Function
Measure or Source
Range
Resolution
Accuracy
Notes
Fluke-719 30G
-12 to 36 psi,
-850 mbar to 2.4 bar
0.001 psi, 0.1 mbar 0.025 % range for 6
months.
0.035 % of range for
one (1) year
Fluke-719 100G
-12 to 120 psi,
-850 mbar to 8 bar
0.01 psi, 1 mbar
mA
0 to 24 mA
0.001 mA
.015 % Rdg + 2 counts Max load, 1000
Ω
7
50
Ω
ohm in HART mode
Loop supply
24 V dc
N/A
±