![Fluke 714B User Manual Download Page 22](http://html1.mh-extra.com/html/fluke/714b/714b_user-manual_2310394022.webp)
714B
Users Manual
16
Table 6. Thermocouple Types Accepted
Type
Positive Lead
Material
Specified Range
(
°
C)
Positive Lead (H) Color
Negative Lead
Material
ANSI* IEC**
E
Chromel
-250 to 1000
Purple
Violet
Constantan
N
Ni-Cr-Si
-200 to 1300
Orange
Pink
Ni-Si-Mg
J
Iron
-210 to 1200
White
Black
Constantan
K
Chromel
-200 to 1372
Yellow
Green
Alumel
T
Copper
-250 to 400
Blue
Brown
Constantan
B
Platinum - 30 % Rhodium
600 to 1820
Gray
Platinum - 6 % Rhodium
R
Platinum - 13 % Rhodium
-20 to 1767
Black
Orange
Platinum
S
Platinum - 10 % Rhodium
-20 to 1767
Black
Orange
Platinum
C
Tungsten – 5 % Rhenium
0 to 2316
White
None
Tungsten - 26 % Rhenium
L Iron
-200 to 900
Constantan
U Copper
-200 to 400
Constantan
BP
90.5 % Ni - 9.5 % Cr
0 to 2500
GOST
56 % Cu - 44 % Ni
Violet or Black
XK
95 % W - 5 % Re
-200 to 800
Red or Pink
80 % W - 20 % Re
www.
GlobalTestSupply
.com
Find Quality Products Online at: