background image

mA Calibrator 

Calibration Adjustment 

Table 4. DC Current Source / Current Measure / Voltage Measure Mode Tests  

Test No. 

707Ex mA Calibrator Output 

Mode 

HP 3458A Minimum 

HP 3458A Maximum 

4.000 mA 

12.000 mA 

24.000  mA 

mA Source 

mA Source 

mA Source 

3.9974 mA 

11.9962 mA 

23.9944 mA 

4.0026 mA 

12.0038 mA 

24.0056 mA 

Test No. 

5520A Calibrator Output 

Mode 

Fluke 707Ex Minimum 

Fluke 707Ex Maximum 

4.000 mA 

12.000 mA 

24.000  mA 

mA Measure 

mA Measure 

mA Measure 

3.997 mA 

11.996 mA 

23.994  mA 

4.003 mA 

12.004 mA 

24.006  mA 

Test No. 

5520A Calibrator Output 

Mode 

Fluke 707Ex Minimum 

Fluke 707Ex Maximum 

0.000 V 

14.000 V 

28.000 V 

Volt Measure 

Volt Measure 

Volt Measure 

-0.002 V 

13.995 V 

27.993 V 

0.002 V 

14.005 V 

28.007  V 

 

DC Voltage Measurement Mode 

Tests 7-9: 

1.

 

Press 

A

 twice to get the mA Calibrator to  

V MEASURE

. The mA Calibrator’s display 

should read "

V

". 

2.

 

Connect the test leads from the 

NORMAL

 output 

terminals of the Fluke 5520A to the input 

terminals on the mA Calibrator (black to 

COM

 

and red to 

+

). 

3.

 

Set the Fluke 5520A to test 7 in Table 4 and 

verify the display reading on the mA Calibrator. 

4.

 

Repeat for tests 8 and 9. 

5.

 

The reading on the display should be within the 

minimum and maximum values shown in  

Table 4. 

6.

 

Set the 5520A output to 0 V (zero), and set output 

to 

STANDBY

7.

 

Turn the mA Calibrator 

OFF

 and disconnect the 

mA Calibrator from the 5520A.  

8.

 

The performance verification tests are now 

complete. 

9.

 

If the mA Calibrator failed any of these tests, 

calibration adjustment or repair is required. 

Calibration Adjustment  

Perform the following calibration adjustment 

procedure if the mA Calibrator fails the performance 

verification test.  

Calibration Adjustment Counter 

The mA Calibrator contains a calibration adjustment 

counter. The value in the calibration adjustment 

counter can be recorded and used to show that no 

adjustments have been made during a calibration 

cycle. When Fluke performs the calibration, the value 

of the calibration adjustment counter will be recorded 

on the calibration label.  

To find the value in the calibration adjustment 

counter, hold down 

B

 

while turning the unit on. 

The software revision will display for 2 seconds 

followed by the calibration counter for another 2 

seconds. 

Note 

Make sure that the mA Calibrator has a new 

battery before starting the calibration 

procedure. Calibration will not function 

properly if the low battery indicator 

(

M

)

 is 

on. 

Because the mA Calibrator incorporates several key 

hold start up features, entering the Adjust mode 

through the keypad requires an exact key hold 

sequence: 

1.

 

With the mA Calibrator off, hold down 

B

 

and 

C

 and turn the mA Calibrator on. 

"

CAL

"

 appears on the main display and the 

calibration adjustment counter is displayed on the 

Summary of Contents for 707Ex

Page 1: ...ete operating instructions refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator Warranty This Fluke product comes with a 3 year warranty for material and workmanship For warranty information refer to the 707Ex Users Manual that came with the mA Calibrator Contacting Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE...

Page 2: ... to 18 C and 28 C to 50 C Relative Humidity 95 up to 30 C 75 up to 40 C 45 up to 50 C Vibration Random 2 g 5 to 500 Hz Shock 1 meter drop test Safety Compliance Complies with IEC 61010 1 95 CAT I 28 V CSA C22 2 No 1010 92 NRTL ANSI ISA S82 02 01 1994 Directive 94 9 EG and p NEC 500 Uo 27 6 V Io 96 13 mA Co 76 nF Lo 2 5 mH Ui 30 V Ii 24 mA Ci 10 nF Li 0 mH CE Complies with EN61010 1 and EN61326 Pow...

Page 3: ...t is used Measurement transients are defined in IEC1010 1 as 2 µs rise time with a 50 µs duration at 50 of the maximum amplitude height Measurement Category I CAT I is defined for measurements performed on circuits not directly connected to the mains Do not use in a damp or wet environment Faults and Damage XW Warning Applying a voltage greater than 28 V to the input of the mA Calibrator invalidat...

Page 4: ...0460 2 P Conforms to European Union requirements F Direct current Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Conforms to relevant Australian standards Maintenance XW Warning To avoid electric shock personal injury or damage to the mA Calibrator Do not service this product To maintain the integrity of the mA Calibrator in explosive atm...

Page 5: ...or handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE 3 DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES USE A HIGH RESIS TANCE GROUNDING WRIST STRAP 4 HANDLE S S DEVICES BY THE BODY static a...

Page 6: ...VE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY ANTI STATIC TYPE SOLDER SUCKERS SHOULD BE USED 11 ONLY GROUNDED TIP SOLDERING IRONS SHOULD BE USED PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC AND GERNER DYNAMICS POMON...

Page 7: ...ery and re install the battery door Make sure it is securely in place 8 Return the mA Calibrator to its red holster aqv07f eps Figure 1 Replacing the Battery User Replaceable Parts User replaceable parts and accessories are shown in Figure 2 and Table 1 1 2 3 6 5 4 7 aqv10f eps Figure 2 User Replaceable Parts Table 1 User Replaceable Parts Item Part Description Fluke Part No Qty A Non skid foot 88...

Page 8: ...C Amps DCI 6 Refer to Table 4 to verify the readings on the HP 3458A for the following tests No adjustment is necessary for test 1 Test 2 1 Press B twice to select 12 000 mA 2 Verify the readings Test 3 1 Press B twice to select 20 000 mA 2 Press and rotate the knob until 24 000 mA is displayed 3 Verify the readings DC Current Measurement Mode Tests 4 6 1 Press the A key twice to put the mA Calibr...

Page 9: ...and set output to STANDBY 7 Turn the mA Calibrator OFF and disconnect the mA Calibrator from the 5520A 8 The performance verification tests are now complete 9 If the mA Calibrator failed any of these tests calibration adjustment or repair is required Calibration Adjustment Perform the following calibration adjustment procedure if the mA Calibrator fails the performance verification test Calibratio...

Page 10: ...st 6 Disconnect the HP 3458A from the mA Calibrator 7 Connect the test leads from the AUX terminals of the Fluke 5520A to the terminals on the mA Calibrator black to COM and red to 8 The mA Calibrator will display MEASURE and 0 000 mA is displayed On the secondary display the last 3 digits of the A D reading are shown Set the 5520A to output 0 000 mA into the mA Calibrator Press C 9 After a moment...

Reviews: