Fluke 700P2X Series Instruction Sheet Download Page 2

15 PSID/G

PRESS

URE M

OD

ULE

70

0P0

4

1 ba

r

10

0 kP

a

RANGE

Figure 1.

Avoiding Overpressure Damage

Applying pressure in excess of the BURST PRESSURE
specified on the pressure module can destroy the pressure
module. Burst pressure is 3X full scale.

Avoiding Corrosion Damage

To avoid damaging the pressure module from corrosion, use
it only with specified media as shown below:

  High: any medium that is compatible with type 316

stainless steel.

 

Low: dry, noncorrosive gasses only.

Recommended Measurement Technique

For best results, it is recommended that the module be
pressurized to full scale and then vented to zero pressure
(atmosphere) prior to zeroing and making measurements.

Note

Low range pressure modules may be sensitive to
gravity. For best results, pressure modules 30 psi
and below should be held at the same physical
orientation from the time they are zeroed until the
measurement is complete.

Pressure Calibration Kit

The Fluke-700PCK Pressure Calibration Kit makes it possible
to calibrate pressure modules at ambient temperature with a
precision pressure calibrator better than the module
specification. A 386 or better PC and Windows

®

 3.1 or later

are required. The kit is an optional accessory available from
your distributor or Fluke.

Performance Test

If you need to check that the pressure module meets its
accuracy specification, use a dead weight tester or suitable
pressure calibrator. The accuracy of the dead weight tester or
pressure calibrator should be significantly better than the
700P2X Series pressure specification. Proceed as follows to
verify that a pressure module is operating within
specification:
1.  Read the pressure value with no externally applied

pressure to make sure the 0% of scale is correct. When
reading the pressure, press the ZERO key to remove
any zero offset.

Note

The pressure ZERO function is available on Fluke-
701 and Fluke-702 calibrators with V1.3 or higher
software. Contact your Fluke Service Center for
upgrade of earlier calibrators.

2.  Connect the pressure module to a dead weight tester.

3.  Set the dead weight tester to 20% of the pressure

module’s full scale value.

4.  Make sure the reading agrees with the dead weight

tester value within the specification in Table 1.

5.  Set the dead weight tester to 40, 60, 80, and 100% of

full scale and compare the respective readings.

6.  If temperature sensitivity is of concern, repeat steps 1

through 5 at various controlled temperatures.

Table 1. Specifications 

(% of full span)

Model

Range

2

Gage or

Differential

Isolated or

Nonisolated

Reference

Uncertainty
(23 

°

 

±

 3 

°

C)

Stability

(1 Year)

Temp

(0 to 50 

°

C)

Total

Uncertainty

3

Fluke-700P22

0 to 1.0000 psi
0 to 6.8900 kPa
0 to 6.89 E-2 bar

Differential

Isolated

0.100%

0.020%

0.030%

0.150%

Fluke-700P23

0 to 5.0000 psi
0 to 34.000 kPa
0 to 0.3400 bar

Differential

Isolated

0.025%

0.010%

0.015%

0.050%

Fluke-700P24

0 to 15.000 psi
0 to 100.00 kPa
0 to 1.0000 bar

Differential

Isolated

0.025%

0.010%

0.015%

0.050%

1. 

Use of pressure zero function is required to achieve these specifications. Contact your Fluke Service Center for upgrade of your Fluke 701 or
Fluke 702 V1.0, V1.1, or V1.2 Calibrator.

2. 

Available pressure units (inHg, kg/cm

2

, mmH

2

0) are determined by the calibrator being used.

3. 

Accuracy specifications apply for 1 year for 0 to 100% of full span from 0 to 50 

°

C. Typical uncertainty is 1% of full span from

 -10 

°

C to 0 

°

C.

4. 

Maximum non-destructive pressure: 3X maximum rated pressure, including common mode pressure.

5. 

Maximum common mode pressure: 3X maximum rated pressure.

6. 

Specifications reflect a confidence interval of 95%.

Reviews: