Fluke 700G01 User Manual Download Page 27

 

Pressure Gauge 

 

Specifications 

19 

 

Model Number 

  

700GA4 

700GA5 

700GA6 

700GA27  700RG05 700RG06 700RG07 700RG08 700RG29 700RG30 700RG31

Pressure 
Range (psi) 

 

15 PSIA 

30 PSIA 

100 PSIA 

300 PSIA 

30 

100 

500 

1000 

3000 

5000 

10000 

Vacuum 
Range (psi) 

 

0 PSIA 

0 PSIA 

0 PSIA 

0 PSIA 

-14 

-12 

-12 

-14 

-14 

-14 

-14 

Burst 
Pressure  
(psi) 

 

60  120 400  1200 90  400 2000 

4000 

10000 

15000 

20000 

Proof 
Pressure (psi) 

 

30  60 200 600 60 200 

1000 

2000 

6000 

10000 

15000 

  

 

 

 

 

 

 

 

 

 

 

 

  

Engineering 
Unit 

Factor 

 

 

 

 

 

 

 

 

 

 

  

mH2O @ 4°C  

0.703089 

10.546 21.093 70.309  210.93 

21.093 70.309 351.54 703.09 2109.3 

3515.4 7030.9 

mH2O @ 20°C  

0.704336 

10.565 21.130 70.434  211.30 

21.130 70.434 352.17 704.34 2113.0 

3521.7 7043.4 

inH2O @ 4°C  

27.68067 

415.21 830.42 2768.1  8304.2 

830.42 2768.1 13840  27681  83042 * 

 

 

inH2O @ 20°C  

27.72977 

415.95 831.89 2773.0  8318.9 

831.89 2773.0 13865  27730  83189 * 

 

 

inH2O @ 60°F  

27.70759 

415.61 831.23 2770.8  8312.3 

831.23 2770.8 13854  27708  83123 * 

 

 

ftH2O @ 4°C 

2.306726 

34.601 69.202 230.67  692.02 

69.202 230.67 1153.4 2306.7 6920.2 

11534  23067 

ftH2O @ 20°C  

2.310814 

34.662 69.324 231.08  693.24 

69.324 231.08 1155.4 2310.8 6932.4 

11554  23108 

ftH2O @ 60°F  

2.308966 

34.634 69.269 230.90  692.69 

69.269 230.90 1154.5 2309.0 6926.9 

11545  23090 

ft Sea Water  

2.24719101 

33.708 67.416 224.72  674.16 

67.416 224.72 1123.6 2247.2 6741.6 

11236  22472 

m Sea Water  

0.68494382 

10.274 20.548 68.494  205.48 

20.548 68.494 342.47 684.94 2054.8 

3424.7 6849.4 

Torr  

51.71507 

775.73 1551.5 5171.5  15515 

1551.5 

5171.5 

25858 

51715 

*  

*  

*  

* - range will not be displayed due to limitations on display resolution. In all cases, resolution is limited to 100,000 counts. 

 

Summary of Contents for 700G01

Page 1: ...Rev 1 10 13 2011 2013 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 700G Series Pressure Gaug...

Page 2: ...ce center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNES...

Page 3: ...Equipment 2 Safety Information 2 Hazard Location Information Approvals 3 Special Conditions for Safe Use 3 Symbols 4 Display and Buttons 5 Operation 7 How to Setup the Product 7 Engineering Units 7 Se...

Page 4: ...Engineering Unit or Scale 9 Battery Life 9 Maintenance 10 How to Clean the Product 10 How to Change the Batteries 10 Accessories 11 RS 232 Interface 11 Specifications 12 Available Input Ranges 12 Acc...

Page 5: ...iii List of Tables Table Title Page 1 Symbols 4 2 Display and Buttons 5...

Page 6: ...700G Series Users Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 The Product 5 2 Change the Batteries 11...

Page 8: ...700G Series Users Manual vi...

Page 9: ...ck its settings and use password protection to prevent configuration changes How to Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 80...

Page 10: ...rdous and the energy from them can be released without warning Use the Product only as specified or the protection supplied by the Product can be compromised The battery door must be closed and locked...

Page 11: ...0014 X Ta 10 C 55 C Special Conditions for Safe Use Misuse If the Product is exposed to overpressure or sudden physical shock such as being dropped examine it for any damage that can cause a safety co...

Page 12: ...th Korean EMC Standards Conforms to relevant Australian standards Conforms to ATEX requirements This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicate...

Page 13: ...e Display and Buttons 5 Display and Buttons The Display and Buttons are shown in Figure 1 The Buttons are explained in Table 2 10K PSI 700 BAR 700G31 10K PSIG PRESSURE GAUGE 3 1 2 4 gsn001 eps Figure...

Page 14: ...he Product on Push again to turn it off Zeros the display In Configure Mode push the button to move forward through the menus Note For absolute versions of the gauge pushing prompts you to enter a bar...

Page 15: ...he gauge goes back to live operation To clear the MIN MAX memory values push and hold for 2 seconds until CLr is shown In Configure Mode push to move backward through the menus Push to go to setup and...

Page 16: ...dard engineering units plus one custom unit scale When the necessary unit shows push or Pressure now shows in the chosen engineering units See the Specifications section for a list of available engine...

Page 17: ...d to the gauge Function Lock When set access to each of the settable parameters above can be turned off to prevent unauthorized configuration changes This is done with password protection in Superviso...

Page 18: ...Custom Engineering Unit or Scale The last menu selection in Supervisory mode is SET FACTR You can set a multiplier factor from 0 001 to 100 to make a custom scale The set factor is multiplied by the...

Page 19: ...amage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent battery leakage Have an approved technician repair the Product How to Change the B...

Page 20: ...k of the Product You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC An RS 232 USB cable is sold separately and includes 700G TRA...

Page 21: ...Positive Pressure 700G01 700G02 0 1 FS Vacuum 0 1 FS Temperature Compensation 15 C to 35 C 59 F to 95 F to rated accuracy Note For temperatures from 10 C to 15 C and 35 C to 55 C add 003 FS C Accurac...

Page 22: ...cturer s specification not to exceed storage specification without batteries Without Batteries 40 C to 70 C 40 F to 158 F Humidity 10 to 95 RH Non condensing Pollution Degree 2 IP Rating 64 with boot...

Page 23: ...3 7 cm 4 5 x 5 in depth 1 5 in Pressure Connection in NPT Male Housing Cast ZNAL Display 5 1 2 Digits 16 53 mm 0 65 in high 20 Segment bar graph 0 to 100 Power Battery three size AA alkaline batterie...

Page 24: ...2068 4 6894 8 20684 34474 68948 kPa 6 894757 2 7579 6 8948 103 42 206 84 689 48 2068 4 3447 4 6894 8 13790 20684 34474 68948 MPa 0 006894757 0 0028 0 0069 0 1034 0 2068 0 6895 2 0684 3 4474 6 8948 13...

Page 25: ...1 55361 83042 inH2O 20 C 27 72977 11 092 27 730 415 95 831 89 2773 0 8318 9 13865 27730 55460 83189 inH2O 60 F 27 70759 11 083 27 708 415 61 831 23 2770 8 8312 3 13854 27708 55415 83123 ftH2O 4 C 2 30...

Page 26: ...68 4 6894 8 20684 2068 4 6894 8 34474 68948 kPa 6 894757 103 42 206 84 689 48 2068 4 206 84 689 48 3447 4 6894 8 20684 34474 68948 MPa 0 006894757 0 1034 0 2068 0 6895 2 0684 0 2068 0 6895 3 4474 6 89...

Page 27: ...1 8304 2 830 42 2768 1 13840 27681 83042 inH2O 20 C 27 72977 415 95 831 89 2773 0 8318 9 831 89 2773 0 13865 27730 83189 inH2O 60 F 27 70759 415 61 831 23 2770 8 8312 3 831 23 2770 8 13854 27708 83123...

Page 28: ...700G Series Users Manual 20...