9-1
Chapter 9
Controller Operation
This section discusses in detail how to operate the bath temperature controller
using the front control panel. Using the front panel key switches and LED display
the user may monitor the bath temperature, set the temperature set-point in
degrees C or F, monitor the heater output power, adjust the controller
proportional band, set the cut-out set-point, and program the probe calibration
parameters, operating parameters, serial and IEEE-488 interface configuration,
and controller calibration parameters. Operation is summarized in Figure 7.
9.1 Bath
Temperature
The digital LED display on the front panel allows direct viewing of the actual bath
temperature. This temperature value is what is normally shown on the display.
The units, C or F, of the temperature value are displayed at the right. For
example,
25.00 C
Bath temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
9.2 Reset Cut-out
If the over-temperature cut-out has been triggered then the temperature display
will alternately flash,
cut-out
Indicates cut-out condition
The message will continue to flash until the temperature is reduced and the cut-
out is reset.
Summary of Contents for 7008
Page 6: ...7008 Users Guide iv ...
Page 8: ...7008 Users Guide vi ...
Page 10: ...7008 Users Guide viii ...
Page 18: ...7008 Users Guide 2 2 Figure 1 Bath Assembly ...
Page 24: ...7008 Users Guide 4 4 ...
Page 45: ...Controller Operation 9 2 Reset Cut out 9 9 3 Figure 7 Controller Operation Flowchart ...
Page 58: ...7008 Users Guide 9 16 ...
Page 62: ...7008 Users Guide 10 4 Table 5 Interface Command Summary ...
Page 70: ...7008 Users Guide 11 4 ...